Chapter 6
141
Custom Arb Waveform Generator
Working with Modulation Types
Creating an FSK Modulation Type User File with the Frequency Values Editor
Use this procedure to set the frequency deviation for data
00
,
01
,
10
, and
11
to configure a user-defined
FSK modulation.
1. Press
Preset
.
2. Press
Mode
>
Custom
>
Real Time I/Q Baseband
>
Modulation Type
>
Define User FSK
>
More (1 of 2)
>
Delete All Rows
>
Confirm Delete All Rows
.
This accesses the
Frequency Values
editor and clears the previous values.
3. Press
600
>
Hz
.
4. Press
1.8
>
kHz
.
5. Press
-600
>
Hz
.
6. Press
-1.8
>
kHz
.
Each time you enter a value, the
Data
column increments to the next binary number, up to a total of 16
data values (from
0000
to
1111
). An unstored file of frequency deviation values is created for the
custom 4-level FSK file.
7. Press
Load/Store
>
Store To File
.
If there is already a file name from the
Catalog of FSK Files
occupying the active entry area, press
the following keys:
Edit Keys
>
Clear Text
8. Enter a file name (for example,
NEWFSK
) using the alpha keys and the numeric keypad.
9. Press
Enter
.
The user-defined FSK modulation should now be stored in the
Catalog of FSK Files
.
Содержание E8247C
Страница 10: ...Contents x ...
Страница 96: ...86 Chapter 4 Analog Modulation Configuring the LF Output ...
Страница 127: ...Chapter 5 117 Dual Arbitrary Waveform Generator Using Waveform Clipping Figure 5 16 Rectangular Clipping ...
Страница 128: ...118 Chapter 5 Dual Arbitrary Waveform Generator Using Waveform Clipping Figure 5 17 Reduction of Peak to Average Power ...
Страница 142: ...132 Chapter 6 Custom Arb Waveform Generator Working with Filters ...
Страница 178: ...168 Chapter 7 Custom Real Time I Q Baseband Working with Differential Data Encoding ...
Страница 198: ...188 Chapter 10 Troubleshooting ...
Страница 214: ...204 Chapter 10 Troubleshooting Returning a Signal Generator to Agilent Technologies ...