Register Definitions
B
E6198B Switch/Load Unit User Manual
B-27
•
Write Only
•
State: 1 = NO shorted to COM, NC is open; 0 = NC shorted to
OM, NO is open
•
Power On/Reset State = 0
Power Select 9-16 (W) Base + 8h
This register controls the Pull Up/Down Power Select armature
relays for channels 9-16. These relays are Form C and the
register uses positive logic: 1 = NO shorted to COM, NC is open;
0 = NC shorted to COM, NO is open. The power buses selected
are determined by how the card is configured. Writing to this
register starts the relay timer.
•
Write Only
•
State: 1 = NO shorted to COM, NC is open; 0 = NC shorted to
COM, NO is open
•
Power On/Reset State = 0
Power Select 17-24 (W) Base + 9h
This register controls the Pull Up/Down Power Select armature
relays for channels 17-24. These relays are Form C and the
register uses positive logic: 1 = NO shorted to COM, NC is open;
0 = NC shorted to COM, NO is open. The power buses selected
are determined by how the card is configured. Writing to this
register starts the relay timer.
Bit
7
6
5
4
3
2
1
0
Purpose
K16
CH16
K15
CH15
K14
CH14
K13
CH13
K12
CH12
K11
CH11
K10
CH10
K9
CH9
Setting
state
state
state
state
state
state
state
state
Bit
7
6
5
4
3
2
1
0
Purpose
K24
CH24
K23
CH23
K22
CH22
K21
CH21
K20
CH20
K19
CH19
K18
CH18
K17
CH17
Setting
state
state
state
state
state
state
state
state
Содержание E6198B
Страница 1: ...Agilent Technologies Agilent TS 5000 Functional Test System E6198B Switch Load Unit User Manual...
Страница 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 17: ...0 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Declaration of Conformity...
Страница 129: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 61 Figure 5 42 Agilent E6178B Load Card Wiring...
Страница 151: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 83 Figure 5 58 Agilent N9378A UUT Connector...
Страница 158: ...5 90 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 188: ...6 30 E6198B Switch Load Unit User Manual 6 Using the Pin Matrix Cards THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 197: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 9 Figure 7 7 E8794A Component Locator...
Страница 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...
Страница 199: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 11 Figure 7 9 Status and Control Interface...
Страница 200: ...7 12 E6198B Switch Load Unit User Manual 7 Using the Custom Card THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 303: ...Repair Information C E6198B Switch Load Unit User Manual C 3 Figure C 1 Agilent E6175A Component Locator Rear View...
Страница 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...
Страница 318: ...G 6 E6198B Switch Load Unit User Manual G Glossary Of Terms THIS PAGE IS INTENTIONALLY LEFT BLANK...