Using the Pin Matrix Cards
6
E6198B Switch/Load Unit User Manual
6-3
•
Automatic disconnecting of column relays for minimal
loading of the Analog Bus,
•
A single control bit can open all relays (OAR),
•
Auxiliary or direct row access relays on each row,
•
Independently switchable series resistance protection on
each column.
Detailed Block Diagram Descriptions
Figure 6-2
is a detailed block diagram of the Agilent E8792A
32-Pin Matrix Card and
Figure 6-3
is a detailed block diagram of
the Agilent E8793A 32-Pin Matrix Card.
Differences Between the Cards
The primary difference between the two cards is that the
Agilent E8792A contains a 16 x 5 Instrument Matrix. The
Instrument Matrix is used to connect measurement or source
instruments to the Analog Bus. Also notice in <Blue>Figure the
DAC1 or DAC2 inputs to the Instrument Matrix. These lines
come from the Switch/Load Units DACs and can be switched
into the Instrument Matrix. The Agilent E8793A also contains
two additional sets of Analog Bus access on connector J1. Other
than these differences, the two cards are electrically and
functionally identical. The following discussion applies to both
cards.
Features Common to Both Cards
The 32-Pin Matrix Modules contain a 32 x 4 matrix of relays,
additional relays to connect/disconnect signals on the buses,
programmable registers to control the relays (described in
Appendix B), and various other features. All relays are of the
high-speed, dry reed type for fast switching.
As shown in
Figure 6-2
and
Figure 6-3
, the Measurement Matrix
is arranged in 32 rows that can be connected to any of four
columns on the common Analog Bus. Closing a matrix relay
connects a row to a column on the card. The columns are
connected to the Analog Bus which carries the signal between
the UUT (unit under test) and instruments connected to the
Analog Bus through the Agilent E8792A. This 32 x 4 structure
lets you connect any system resource to any pin on the UUT.
This matrix along with the unswitched UUT Common allows as
many as four system resources to be connected simultaneously.
Содержание E6198B
Страница 1: ...Agilent Technologies Agilent TS 5000 Functional Test System E6198B Switch Load Unit User Manual...
Страница 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 17: ...0 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Declaration of Conformity...
Страница 129: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 61 Figure 5 42 Agilent E6178B Load Card Wiring...
Страница 151: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 83 Figure 5 58 Agilent N9378A UUT Connector...
Страница 158: ...5 90 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 188: ...6 30 E6198B Switch Load Unit User Manual 6 Using the Pin Matrix Cards THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 197: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 9 Figure 7 7 E8794A Component Locator...
Страница 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...
Страница 199: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 11 Figure 7 9 Status and Control Interface...
Страница 200: ...7 12 E6198B Switch Load Unit User Manual 7 Using the Custom Card THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 303: ...Repair Information C E6198B Switch Load Unit User Manual C 3 Figure C 1 Agilent E6175A Component Locator Rear View...
Страница 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...
Страница 318: ...G 6 E6198B Switch Load Unit User Manual G Glossary Of Terms THIS PAGE IS INTENTIONALLY LEFT BLANK...