Using the Pin Matrix Cards
6
E6198B Switch/Load Unit User Manual
6-7
Relay Timer
The Relay Timer indicates whether or not the last change to a
relay's state (opening or closing the relay) is complete. The
timer starts or restarts when a command to change a relay's
state is received. The timer stops when it times out after an
interval sufficiently long for a relay to change state.
If the relay timer's status is “busy,” as reported by bit 7 of the
Status Register (see
Appendix B
), relays may not yet be in the
desired state. If the status is “not busy,” then the relays have
reached their newly programmed state. The nominal time-out
value of the relay counter is 500
μ
s.
Column Disconnect Relay Control
Each of the four columns has a disconnect relay between it and
the Analog Bus. Depending on the state of bit 6 in the Control
Register (described in
Appendix B
), the disconnect relays are
either under manual or automatic control.
When in automatic mode, when you close a matrix relay, the
disconnect relay associated with that column also closes. When
in manual mode, the four column disconnect relays are
controlled by bits 3-0 in the Column Control & Protection
Bypass Register (described in
Appendix B
). Because manual
mode is used only when doing diagnostic checks, the default
control mode is automatic.
OAR
Bit 5 in the Control Register controls the OAR (“open all relays”)
feature, which immediately opens all relays on the card.
Because the bit is self-clearing, it does not require resetting.
OAR also clears bits 3-0 of the Column Control & Protection
Bypass Register. When executed, OAR re-triggers the relay
timer.
Reset
Bit 0 in the Control Register programmatically resets the card.
Resetting the card clears all internal registers, which resets all
board functionality to its default, power-up state. Resetting also
clears all relay registers, sets column disconnect relay control to
automatic mode, and starts the relay timer.
Содержание E6198B
Страница 1: ...Agilent Technologies Agilent TS 5000 Functional Test System E6198B Switch Load Unit User Manual...
Страница 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 17: ...0 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Declaration of Conformity...
Страница 129: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 61 Figure 5 42 Agilent E6178B Load Card Wiring...
Страница 151: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 83 Figure 5 58 Agilent N9378A UUT Connector...
Страница 158: ...5 90 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 188: ...6 30 E6198B Switch Load Unit User Manual 6 Using the Pin Matrix Cards THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 197: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 9 Figure 7 7 E8794A Component Locator...
Страница 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...
Страница 199: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 11 Figure 7 9 Status and Control Interface...
Страница 200: ...7 12 E6198B Switch Load Unit User Manual 7 Using the Custom Card THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 303: ...Repair Information C E6198B Switch Load Unit User Manual C 3 Figure C 1 Agilent E6175A Component Locator Rear View...
Страница 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...
Страница 318: ...G 6 E6198B Switch Load Unit User Manual G Glossary Of Terms THIS PAGE IS INTENTIONALLY LEFT BLANK...