Using the Pin Matrix Cards
6
E6198B Switch/Load Unit User Manual
6-21
Relay Timer
The Relay Timer indicates whether or not the last change to a
relay's state (opening or closing the relay) is complete. The
timer starts or restarts when a command to change a relay's
state is received. The timer stops when it times out after an
interval sufficiently long for a relay to change state. If the relay
timer's status is "busy," as reported by bit 7 of the Status
Register (see
Appendix B
), relays may not yet be in the desired
state. If the status is "not busy," then the relays have reached
their newly programmed state. The nominal time-out value of
the relay counter is 500 us.
OAR
Bit 5 in the Control Register controls the OAR ("open all relays")
feature, which immediately opens all relays on the card.
Because the bit is self-clearing, it does not require resetting.
OAR also clears bits 3-0 of the Column Control & Protection
Bypass Register. When executed, OAR re-triggers the relay
timer.
Reset
Bit 0 in the Control Register programmatically resets the card.
Resetting the card clears all internal registers, which resets all
board functionality to its default, power-up state. Resetting also
clears all relay registers and starts the relay timer.
Protection Bypass
Each of the four columns has a 200 ohm protection resistor
connected in series which protects the relays by limiting the
maximum current through the column. Some measurements
(such as 2-wire resistance) may require bypassing (shorting
across) the protection resistor to remove its effects. The default
state is to have the protection bypass relays open, which means
the series protection resistors are in circuit. Bits 3-0 in the
ABUS Control & Protection Bypass Register (described in
Appendix B
) control the relays used to bypass the series
protection resistors. You can bypass the protection on a
column-by-column basis.
CAUTION
To prevent damage to card, bypas the protection resistors only
when absolutely necessary.
Содержание E6198B
Страница 1: ...Agilent Technologies Agilent TS 5000 Functional Test System E6198B Switch Load Unit User Manual...
Страница 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 17: ...0 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Declaration of Conformity...
Страница 129: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 61 Figure 5 42 Agilent E6178B Load Card Wiring...
Страница 151: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 83 Figure 5 58 Agilent N9378A UUT Connector...
Страница 158: ...5 90 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 188: ...6 30 E6198B Switch Load Unit User Manual 6 Using the Pin Matrix Cards THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 197: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 9 Figure 7 7 E8794A Component Locator...
Страница 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...
Страница 199: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 11 Figure 7 9 Status and Control Interface...
Страница 200: ...7 12 E6198B Switch Load Unit User Manual 7 Using the Custom Card THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 303: ...Repair Information C E6198B Switch Load Unit User Manual C 3 Figure C 1 Agilent E6175A Component Locator Rear View...
Страница 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...
Страница 318: ...G 6 E6198B Switch Load Unit User Manual G Glossary Of Terms THIS PAGE IS INTENTIONALLY LEFT BLANK...