Switch/Load Unit and Plug-In Cards

3

E6198B Switch/Load Unit User Manual

3-23

 

 

Fuse Ratings

Fuses are incorporated inside E6198B Switch/Load Unit. There 
are 10 fuses and 7 fuses for Standalone SLU and Integrated SLU 
respectively. Please refer 

Table 3-11 

for more detail.

DC Power Bus Connections

PB1 to PB4 are the DC power bus connections as seen in 

Figure 3-11

 for the standalone unit. Notice the protective cable 

enclosure covering both the power bus cabling and power sense 
cabling for better protection.

For the integrated unit, there is no protective cable enclosure, 
and the cabling would need to be configured by the user directly 
onto the backplane, behind the SLU backcover.

Table 3-11    

Fuses Ratings 

Standalone

Integrated

Description

Fuse1

Fuse1

1A, -12V

Fuse2

Fuse2

6A, +12V_1

Fuse3

Fuse3

6A, +12V_2

Fuse4

Fuse4

Not Used

Fuse5

Fuse5

15A, +5V_1

Fuse6

Fuse6

15A, +5V_2

Fuse7

Fuse7

Not Used

Fuse8

-

30A, PB2

Fuse9

-

30A, PB3

Fuse10

-

30A, PB4

Содержание E6198B

Страница 1: ...Agilent Technologies Agilent TS 5000 Functional Test System E6198B Switch Load Unit User Manual...

Страница 2: ...f Software is subject to Agilent Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will receive no greater than Restricted Rights as defined in...

Страница 3: ...ound The System 2 4 Fuses 2 4 Operator Safety Information 2 4 Safety Symbols and Regulatory Markings 2 5 End of Life Waste Electrical and Electronic Equipment WEEE Directive 2002 96 EC 2 7 3 Switch Lo...

Страница 4: ...Recommendations 4 2 Load Card Types and IDS 4 4 Connecting an Additional Spare Power Supply 4 5 Configuring the Power Busses 4 6 Setting the UUT Power Supply Remote Local Sense Jumpers 4 8 Configuring...

Страница 5: ...77A 24 Channel Load Card 5 38 Card Layout 5 40 Selecting a Power Supply Configuration 5 41 Using the Power Switches as General Purpose Relays 5 41 Connecting Loads 5 42 Load Wiring 5 44 UUT Connection...

Страница 6: ...Installing and Removing Mezzanine Cards 5 81 Wiring the Mezzanine Cards 5 82 UUT Connections 5 82 Using the Agilent N9379A 48 Channel High Density Load Card 5 84 Card Layout 5 86 Load Circuit Details...

Страница 7: ...7 4 J2 Connector Breakouts 7 4 J3 J4 Connector Breakouts DAC 7 5 J3 or J4 Connector Breakouts Event Detector 7 6 J5 and J6 Connector Breakouts 32 Pin Matrix Cards 7 7 E8794A Components and Schematics...

Страница 8: ...4 Card Type R Base 0h B 5 Card Configuration R Base 1h B 5 Status Register R Base 2h B 5 Fixture ID R Base 3h B 6 Digital Input R Base 4h B 6 DAC1 Output MSB W Base 8h B 6 DAC1 Output LSB W Base 9h B...

Страница 9: ...ement Matrix Card B 65 Agilent E8783A 64 Pin Matrix Card B 74 Agilent E8794A Custom Card Register Definitions B 83 C Repair Information Support Strategy C 2 Locating Load Card Components C 2 Agilent E...

Страница 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 11: ...guring the Switch Load Unit Figure 4 1 Example of Load Matrix Custom Cards Loading Order 4 4 Figure 4 2 Component Location Schematic for User Installed External Power Supply 4 5 Figure 4 3 Spare Suppl...

Страница 12: ...etail 5 30 Figure 5 21 Typical Agilent E6176A Load Card Flyback Protection Circuit 5 31 Figure 5 22 Agilent E6176A Connectors J1 J2 and Mating Connectors P1 P2 5 32 Figure 5 23 Agilent E6176A Load Wir...

Страница 13: ...9A 48 Channel High Density Load Card Block Diagram 5 85 Figure 5 60 Agilent N9379A Layout 5 86 Figure 5 61 Load Circuit Details 5 87 Figure 5 62 Agilent N9379A Mezzanine Card 5 88 6 Using the Pin Matr...

Страница 14: ...7 Figure 7 6 E8794A Layout 7 8 Figure 7 7 E8794A Component Locator 7 9 Figure 7 8 E8794A Schematic 7 10 Figure 7 9 Status and Control Interface 7 11 A Switch Load Unit And Card Specifications B Regist...

Страница 15: ...e 3 6 J102 Signal Definitions 3 18 Table 3 7 Switch Load Unit J1 Pinout 3 19 Table 3 8 Switch Load Unit J2 Pinout 3 19 Table 3 9 Switch Load Unit J3 Pinout 3 19 Table 3 10 Power Supply Input for each...

Страница 16: ...Table C 2 Instrument Matrix Relays C 7 Table C 3 Bypass and Disconnect Relays C 7 Table C 4 Miscellaneous Relays C 7 Table C 5 Measurement Matrix Relays for E8782A C 8 Table C 6 Measurement Matrix Re...

Страница 17: ...0 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Declaration of Conformity...

Страница 18: ...for this product is available on the Agilent Technologies website You can search for the DoC by its product model or description at the following web address http regulations corporate agilent com DoC...

Страница 19: ...5000 E6198B Switch Load Unit User Manual Agilent Technologies 1 Legal Information Warranty 1 2 Technology Licenses 1 2 Restricted Rights Legend 1 2 Service And Support 1 3 Agilent On The Web 1 3 Agile...

Страница 20: ...n the separate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accordance with the t...

Страница 21: ...Double click the link to Test Measurement Select your country from the drop down menus The Web page that appears next has contact information specific for your country Agilent By Phone If you do not...

Страница 22: ...2 404 93 40 Denmark 45 70 13 15 15 Finland 358 0 10 855 2100 France 0825 010 700 0 125 minute Germany 07031 464 6333 0 14 minute Ireland 1890 924 204 Israel 972 3 9288 504 544 Italy 39 02 92 60 8484...

Страница 23: ...fety Information 2 2 Safety Summary 2 2 Safety Notice 2 2 General 2 2 Environmental Conditions 2 3 Before Applying Power 2 3 Ground The System 2 4 Fuses 2 4 Operator Safety Information 2 4 Safety Symb...

Страница 24: ...may be impaired if it is used in a manner not specified in the operation instructions CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that...

Страница 25: ...t covers Component replacement and internal adjustments must be made only by qualified service personnel Products that appear damaged or defective should be made inoperative and secured against uninte...

Страница 26: ...t use repaired fuses or short circuited fuse holders To do so could cause a shock or fire hazard For AC inlet fuse use only 5A 250VAC fuse with Time Lag and Low Breaking capacity For other fuse rating...

Страница 27: ...ctrical power distribution Safety Symbols and Regulatory Markings Symbols and markings on the system in manuals and on instruments alert you to potential risks provide information about conditions and...

Страница 28: ...SA mark with the indicators C and US means that the product is certified for both the U S and Canadian markets to the applicable American and Canadian standards The C tick mark is a registered tradema...

Страница 29: ...label see below indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE directive Annex 1...

Страница 30: ...2 8 E6198B Switch Load Unit User Manual 2 Safety and Regulatory Information THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 31: ...6198B Standalone Switch Load Unit Description 3 3 E6198B Integrated Switch Load Unit Description 3 9 Backplane And Breakout Board Connectors and LEDs 3 10 Differentiating E6198B Standalone Option and...

Страница 32: ...Switch Load Unit Rating Electrical Mechanical Table 3 1 Switch Load Unit Rating Electrical Parameter Value Input Voltage 100 240 VAC nominal Frequency 50 60 Hz Power 325 W maximum Table 3 2 Switch Lo...

Страница 33: ...177A 24 channel load card with current sense Agilent E6178B 8 channel high power load card Agilent N9377A 16 channel dual load load card Agilent N9378A 24 channel low resistance load card Agilent N937...

Страница 34: ...In Cards In addition to holding load cards pin matrix cards and custom cards the Switch Load Unit also provides the following capabilities Figure 3 1 Agilent E6198B Standalone Switch Load Unit Front V...

Страница 35: ...of digital input 8 bits of digital output Open Drain Out and 8 bits of TTL level digital output Spare Digital Out There is no handshaking capability in the digital I O Typical usage of the digital I O...

Страница 36: ...14 bit channels of DAC provide 16 volts at 10 mA each To provide this voltage swing a charge pump is used to step up the 12 volt supply The DAC channels are typically used for controlling differentia...

Страница 37: ...upplies connected to power buses 1 4 This sensing compensates the losses in the system wiring to ensure the set voltage is applied to the sense point Switch Load Unit Block Diagram Figure 3 5 shows a...

Страница 38: ...Card Agilent E6198B Switch Load Unit Load Card n Pin Matrix Card n Digital In Power Bus Sense Isense Switch Load Unit Circuitry Digital Out DAC 12V 12V Power Bus Power Supply 1 Power Supply 2 Power Su...

Страница 39: ...The Agilent E6198B now comes as a system integrated unit or a standalone unit The features and functionality remains the same as the previous E6198A Differences between the two are in the cabling and...

Страница 40: ...ibed on the following pages See Table 3 3 for more detail Figure 3 7 Switch Load Unit Backplane Connectors USB port adaptor board Not Shown BACKPLANE SLU T4 T8 T9 T1 T5 T10 T11 T2 T6 T12 T13 T3 T7 T14...

Страница 41: ...one option and is easily accessible from the SLU rear Figure 3 8 shows connector locations on the breakout board external board Refer Table 3 5 for the detailed descriptions Figure 3 8 Switch Load Uni...

Страница 42: ...ty Connector J104 External Reset reserved J105 5V Indicator 5V Status J106 12V Indicator 12V Status J107 12V Indicator 12V Status J108 READY Indicator Control Status J109 12V Output For Fan J110 12V O...

Страница 43: ...tension connector J_3 Utility Connector J102 internal connector J_5 Frame Select Spare Supply Connect to J1 of backplane internal connector J_6 Power Bus Sense Local Remote Connect to J3 of backplane...

Страница 44: ...and 16 21 These busses may be connected together to form a single bus or split into independent current sense busses using Switches S3A S3F on breakout board Standalone are shown in their factory def...

Страница 45: ...EDs are present on the frontpanel of the SLU Frame LEDs The Frame Access LED flashes to indicate a data access has occurred to that Switch Load Unit The five Slot Address LEDs indicate the slot curren...

Страница 46: ...0 Digital In 2 Digital In 4 Digital In 6 Spare_DigOut 0 Spare_DigOut 6 Spare_DigOut 4 Spare_DigOut 2 ISense 1 ISense 4 ISense 3 ISense 2 Power Bus Sense 1 Power Bus Sense 3 System Gnd System Gnd Syst...

Страница 47: ...nection 12 Vdc Supply 12 Vdc Supply 12 Vdc Supply 3 Open Drain Out 7 Open Drain Out 5 Open Drain Out 1 Open Drain Out 3 DAC2 DAC1 No Connection System Gnd System Gnd Spare Supply 12 Vdc Supply 12 Vdc...

Страница 48: ...D for each mass interconnect fixture and read it back ID sent as TTL level bits Isense 1 4 Isense 1 4 These lines connect to the current sense bus on the Switch Load Unit backplane These lines are use...

Страница 49: ...SENSE Bussed Split Default setting for J2 is ISENSE Bussed J3 Pinout Table 3 9 shows pinouts of Power Bus Sense Local Remote Local 1 is selected if LOC_S1 is connected with PB1 Remote 1 is selected if...

Страница 50: ...T14 provides the required 5 volts and 12 volts for powering the backplane and Load Cards Table 3 10 shows each jumpers contains difference power supply Table 3 10 Power Supply Input for each Jumpers...

Страница 51: ...tandalone unit the backplane connectors are extended to the breakout board attached to the SLU rear cover At SLU rear you will find toggle switches for SLU configuration power and sense buses connecto...

Страница 52: ...3 22 E6198B Switch Load Unit User Manual 3 Switch Load Unit and Plug In Cards Figure 3 12 Switch Load Unit Rear View System Integrated Option...

Страница 53: ...r the standalone unit Notice the protective cable enclosure covering both the power bus cabling and power sense cabling for better protection For the integrated unit there is no protective cable enclo...

Страница 54: ...3 24 E6198B Switch Load Unit User Manual 3 Switch Load Unit and Plug In Cards THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 55: ...Card Location Recommendations 4 2 Connecting an Additional Spare Power Supply 4 5 Configuring the Power Busses 4 6 Connecting E6198B to the Computer via USB Interface or Parallel Port 4 12 Adding a S...

Страница 56: ...rds using the following rules Agilent N9379A 48 Channel Load Cards Install any N9379A cards first starting at slot 1 Agilent E6177A 24 Channel Load Cards Install any E6177A cards next Agilent U7177A 2...

Страница 57: ...nt E6178B 8 channel Heavy Duty cards require 2 slots each These cards are installed in even slot locations only Install E6178B cards starting with the first available even slot following the 16 channe...

Страница 58: ...ard Types and IDS Each card is assigned a different type and has a 10 pin connector that lets you assign a unique binary code ID number to each card See Load Card Type and Configuration ID for more in...

Страница 59: ...cy noise on the supply line The auxiliary supply output Spare Supply appears on Switch Load Unit connector J102 pin 79 and on the Test System Interface TC2 pin N4 For SLU Standalone option you can con...

Страница 60: ...ussed together on the PC board providing a common connection for up to three supplies The following examples A and B show the two most common ways of configuring the power busses Example A shows three...

Страница 61: ...used if for example one supply requires local sensing while the other requires remote sensing at the UUT For the standalone SLU users can directly connect to the external connector DC Power Bus Connec...

Страница 62: ...ross the cables between the load and power supply In remote sense the voltage across the load equals the selected power supply voltage In this configuration the power supply automatically increases th...

Страница 63: ...Sense Remote Local swithes to select either Remote or Local mode See Figure 4 9 Figure 4 9 PB Sense Remote Local Interface Standalone SLU For Integrated use Jumper to select the mode on PB Sense Remot...

Страница 64: ...rrent sense readings can be made at a time The jumpers are located into J2 connector Figure 3 7 Removing any jumper splits the current sense bus at that point Cards to the left of the removed jumper s...

Страница 65: ...ntegrated SLU jumpers are used to select between Bussed Split see Figure 4 12 Figure 4 12 ISense Bussed Split Interface Integrated SLU NOTE Ensure Load Cards that will be accessed simultaneously for c...

Страница 66: ...to the Computer via USB Interface or Parallel Port USB interface is introduced in E6198B on top of parallel port interface which is available on the older version on SLU The USB interface connects to...

Страница 67: ...he Frame Select toggle switches sets the address of each Switch Load Unit see Figure 4 14 The toggle switches can be set for any address from zero to seven However for the Integrated SLU jumpers must...

Страница 68: ...nd secure to the rack using 4 spacers 0380 0317 and 4 dress screws 0570 1272 Spacers are only required for the first loadbox to clear the door latch Check the build list for the number of load boxes E...

Страница 69: ...l Load Card 5 25 Using the Agilent E6177A 24 Channel Load Card 5 38 Using the Agilent U7177A 24 Channel Load Card 5 46 Using the Agilent E6178B 8 Channel Load Card 5 54 Using the Agilent N9377A 16 Cha...

Страница 70: ...dium Current Load Card for non inductive loads is intended to be used with loads mounted on the card A 9 inch by 4 inch area of sheet metal is open on the front of the card for mounting small loads Th...

Страница 71: ...intended to be used with externally mounted loads This card is one slot wide and capable of up to 2A continuous carry current The Agilent E6178B 8 Channel Heavy Duty Load Card is designed for very hig...

Страница 72: ...conceptual block diagram of a typical load application having these four main areas Power Supply Load Card Loads Unit Under Test UUT Figure 5 1 Loads Sources Conceptual Overview Unit Under Test V GND...

Страница 73: ...24 Channel Low Resistance Card Type is 5 Agilent N9379A 48 Channel High Density Card Type is 6 Agilent N9377A 16 Channel Dual Load Card Type is 7 Except for the N9378A and N9379A each load card is eq...

Страница 74: ...n the N9379A load card The configuration pins are normally high with a 10 K pull up resistor resident on the load card producing a value of FFh Grounding selected pins creates binary codes which can b...

Страница 75: ...1 4 of the Agilent E6198B Switch Load Unit enclosure closest to the incoming power buses Card Location Recommendations The card type is 01h Load Card Type and Configuration ID The factory default is...

Страница 76: ...5 8 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads Figure 5 3 Agilent E6175A 8 Channel High Current Load Card Block Diagram...

Страница 77: ...n and an NC normally closed terminal Each NO terminal is connected to a pin on J3 and each NC terminal is connected to a pin on J4 Each pin on J3 or J4 can be connected to any of the four power bus li...

Страница 78: ...cer from LEM Inc Figure 5 5 shows the relationship between the sense resistor Rs and the LEM module current transducer only one or the other is used never both Sense Resistors The card comes factory l...

Страница 79: ...load card s printed circuit board Installing a current transducer involves both elements of a channel pair For example if the LEM module were to be installed across channels 1 2 1 Remove the 0 05 cur...

Страница 80: ...s 5 Install two bypass capacitors C41 and C42 both 0 01 F 6 Install the appropriate primary coil range jumpers See Figure 5 7 NOTE Use a wire gauge appropriate for the current through the load Figure...

Страница 81: ...rimary for Various Current Ranges Table 5 2 Components involved in LEM Module Installation Channel s Isense Resistor Rmvd Measuring Resistor Shorting Jumper Bypass Capacitors 0 01 f LEM Module LEM Tap...

Страница 82: ...els 1 and 2 on the component locator diagram of the 8 Channel Load Card It shows the location and polarity orientation for channel 1 s RT1 and RT2 and channel 2 s RT3 and RT4 flyback protection device...

Страница 83: ...s should be installed with the positive side towards the UUT On each of the four channel pairs the high side should be located as shown in the component locator diagram Figure 5 9 Figure 5 8 8 Channel...

Страница 84: ...l MOV 3mm axial lead mounted specifications are Continuous DC voltage 220 V Transient energy 10 1000mS 0 90 Joules watt seconds Transient peak current 8 20mS 100 Amperes Varistor voltage 1 0mA DC 300...

Страница 85: ...sed at the external DMM when the corresponding Isense MUX relays are closed The ground return in this example is assumed to be switched through another load card or connected directly to the UUT In th...

Страница 86: ...ce Relay Timer Relay Slot Decode Logic J2 Configuration Connector Current Sense Control Address Data Power 5 12V P1 J3 NO PB1 PB2 PB3 PB4 PS 1 PS 2 PS 3 NC J5 J4 Load ISense 1 ISense 2 NO PwrSw3 NC RS...

Страница 87: ...ies are not used The power for the bridge is supplied by the module the connection to the UUT power supply is effectively bypassed See Figure 5 12 Use the following table to determine the appropriate...

Страница 88: ...ds are mounted on the Agilent E6175A s sheet metal mounting area The loads are wired to connector P1 which mates to the Agilent E6175A s J1 connector Figure 5 13 shows the Agilent E6175A s load mounti...

Страница 89: ...P1 Connector Wiring Terminal View Row 15 Row 14 Row 13 Row 12 Row 11 Row 10 Row 9 Row 8 Row 7 Row 6 Row 5 Row 4 Row 3 Row 2 Row 1 Channel 1 Power 1 Channel 2 Channel 3 Channel 4 Channel 5 Channel 6 Ch...

Страница 90: ...nected together on the PC board When connecting high current 3 amp loads wire across all three pins in each row of P1 see Figure 5 15 This ensures current sharing across all pins and prevents prematur...

Страница 91: ...ils You can also make connections directly to P2 Figure 5 16 is a P2 connector pinout showing the details Each even numbered row of P2 represents a single load card channel The three pins in each row...

Страница 92: ...d Loads Figure 5 16 Agilent E6175A 8 Channel High Current Load Card UUT Connections NC No Connection P2 Connector Front View E1 NC Channel 8 NC NC NC NC NC NC NC Channel 7 Channel 6 Channel 5 Channel...

Страница 93: ...t required The card is one slot wide Card Location Recommendations The card type is 02h Load Card Type and Configuration ID The factory default configuration is explained in Selecting a Power Supply C...

Страница 94: ...5 26 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads Figure 5 17 Agilent E6176A 16 Channel High Current Load Card Block Diagram...

Страница 95: ...n J4 J7 and each NC terminal is connected to a pin on J8 J11 Each pin on J8 J11 can be connected to any of the four power bus lines on J12 via jumper wires See Figure 5 27 The terminal block jumper al...

Страница 96: ...e sense resistor with a four wire resistor of higher value Selecting a Load Fuse The load fuses used are IEC 5x20 millimeter 5 amp slow blow These fuses may be replaced by fuses with a higher value bu...

Страница 97: ...power switch and one flyback protection from the output to the normally closed side of the power switch In Figure 5 41 Channel 1 has two input lines connected to the input Form C switch RT1 connects...

Страница 98: ...e protection Protection Devices Zener Diodes MOVs Metal Oxide Varistor or Transzorbs devices mounted at RTx or RTy 2 required per load 1 at NC and 1 at NO provide current path for the inductive load f...

Страница 99: ...20mS 450 volts Capacitance 17 pF Typical Transzorb specifications are Breakdown voltage 300V max at 1mA Reverse Standoff voltage 250 volts Maximum reverse leakage 5mA Maximum peak pulse current 5A Ma...

Страница 100: ...e connectors Load Wiring Figure 5 23 is a simplified schematic and connector pinouts showing how loads are connected to P1 P2 Loads 1 through 8 connect to P1 and loads 9 through 16 connect to P2 Figur...

Страница 101: ...nel 5 Channel 6 Channel 7 Channel 8 Power 2 Power 3 Power 4 Power 5 Power 6 Power 7 Power 8 Load 1 Load 2 Load 3 Load 4 Load 5 Load 6 Load 7 Load 8 Row 16 A C E P2 Connector Wiring Terminal View Row 1...

Страница 102: ...ach row of P1 P2 see Figure 5 24 This ensures current sharing across all pins and prevents premature pin failure from excessive current flow CAUTION Connecting high current 4 amp loads without wiring...

Страница 103: ...oose to mount the loads externally Mount the loads as close to the load card as possible The dimensions of the load plate for the 16 channel card are Height 23 32 cm 9 18 in Length Max 34 00 cm 13 386...

Страница 104: ...nnect wiring guide for connection details You can also make connections directly to P2 Figure 5 26 is a simplified load card schematic and P2 connector pinout showing the details Each row of P2 repres...

Страница 105: ...lent E6176A P2 Connector Pinouts P2 Connector Front View E1 Channel 16 Channel 15 Channel 14 Channel 12 Channel 10 Channel 8 Channel 6 Channel 4 Channel 2 Channel 13 Channel 11 Channel 9 Channel 7 Cha...

Страница 106: ...ng a Power Supply Configuration on page 41 The Common line PwrX where X is the channel number on the input is brought back out to the input allowing each channel to operate in a general purpose GP con...

Страница 107: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 39 Figure 5 27 Agilent E6177A 24 Channel Medium Current Load Card Block Diagram...

Страница 108: ...5 40 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads Card Layout Figure 5 28 shows the E6177A layout Figure 5 28 Agilent E6177A 24 Channel Load Card Layout...

Страница 109: ...together and one that ties all the pins on J10 J15 together The NO pins are jumpered to power bus 2 and the NC pins are jumpered to power bus 1 The jumper combs for J3 J8 and J10 Jl5 can be easily cu...

Страница 110: ...re mounted on the Agilent E6177A s sheet metal mounting area The loads are wired to connector P1 which mates to the Agilent E6177A s J1 connector Figure 5 30 shows the Agilent E6177A s load mounting a...

Страница 111: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 43 Figure 5 30 Agilent E6177A Load Mounting Area and P1 J1 Connectors J1 P1 Load Mounting Area...

Страница 112: ...9 C 10 C Channel 11 Load 12 Power 12 9 E 10 E Channel 12 Load 13 Power 13 7 A 8 A Channel 13 Load 14 Power 14 7 C 8 C Channel 14 Load 15 Power 15 7 E 8 E Channel 15 Load 16 Power 16 5 A 6 A Channel 16...

Страница 113: ...ables to a mass interconnect panel User connections to the UUT are then made from the mass interconnect panel Refer to the appropriate mass interconnect wiring guide for connection details You can als...

Страница 114: ...s explained in Selecting a Power Supply Configuration The factory default is to load a 0 05 0 1 sense resistor in each channel See Selecting a Current Sense Resistor Value The Common line PwrX where X...

Страница 115: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 47 Figure 5 33 Agilent U7177A 24 Channel Medium Current Load Card Block Diagram...

Страница 116: ...NC terminal is connected to a pin on J10 J15 Each pin on J3 J15 can be connected to any of the four power busses on J9 via jumper wires See Figure 5 33 The terminal block jumper allows convenient pul...

Страница 117: ...ake the sense voltage readable then replace the sense resistor with a four wire resistor of higher value Using the Power Switches as General Purpose Relays The U7177A has a special feature that is sim...

Страница 118: ...e loads are wired to connector P1 which mates to the Agilent U7177A s J1 connector Figure 5 36 shows these connectors Figure 5 35 Using the 24 Channel Load Card Switches as GP Relays CAUTION When usin...

Страница 119: ...ads 5 E6198B Switch Load Unit User Manual 5 51 Load Wiring Figure 5 37 is a simplified schematic and P1 connector pinout showing how loads are connected to P1 Figure 5 36 Agilent U7177A Connector J1 a...

Страница 120: ...5 52 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads Figure 5 37 Agilent U7177A Load Wiring Schematic and P1 Pinouts...

Страница 121: ...ted via cables to a mass interconnect panel User connections to the UUT are then made from the mass interconnect panel Refer to the appropriate mass interconnect wiring guide for connection details Yo...

Страница 122: ...the external power supplies to the load card not through the Agilent E6198B Switch Load Unit Power Supply Buses Each channel is capable of up to 30 Amps continuous carry into a resistive load Selecti...

Страница 123: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 55 Figure 5 39 Agilent E6178B 8 Channel Heavy Duty Load Card Block Diagram...

Страница 124: ...a load that is floating with respect to the power system Other configurations are possible For example the negative terminals of two external power supplies could be connected together at the UUT and...

Страница 125: ...alled load fuses are x 1 30 amp slow blow These fuses may be replaced by fuses with a lower value depending on your load requirements The Agilent E6178 was qualified with a Bussman MDL 30 fuse This fu...

Страница 126: ...m over the temperature range Offset Error 0 3 Amps Typical over the temperature range 0 7 Amps Maximum over the temperature range The error due to the offset component can be reduced by measuring the...

Страница 127: ...flyback Select the protection device so that it conducts at a voltage higher than the UUT s internal protection If the device s internal protection fails then the added external protection conducts to...

Страница 128: ...2 shows how the Agilent E6178B Load Card is wired Connect loads between J11 J18 and J31 J38 Loads are external to the load card and connected to the load card via cables Connectors J11 and J31 connect...

Страница 129: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 61 Figure 5 42 Agilent E6178B Load Card Wiring...

Страница 130: ...include The card is one slot wide Card Location Recommendations The card type is 07h Load Card Type and Configuration ID Each channel is capable of up to 7 5 amp continuous carry into a resistive load...

Страница 131: ...PB1 PwrSw1 J3 Configuration Connector Relay Timer Relay Slot Decode Logic Protection Mother Board Interface 180 180 UUT Loads P2 Load Switches External Mount Loads NO1 RS NC1 LdSense 1 LdSense 1 Flyb...

Страница 132: ...1 K30 3 K30 2 K30 4 K30 5 K30 7 K30 6 K32 5 K32 4 K32 6 K32 7 K32 9 K32 8 K33 1 K33 0 K31 7 K31 6 K31 9 K31 8 K32 0 K32 1 K32 2 K32 3 Relays Sense elements flyback clamps Fuses Reed relays Isense Rel...

Страница 133: ...on J8 J11 NC together The NO pins are jumpered to power bus 2 and the NC pins are jumpered to power bus 1 The jumper combs for J4 Jll can be easily cut to provide bus or pin isolation between the var...

Страница 134: ...and channel 2 RT503 and RT504 flyback protection devices when they are installed This pattern is repeated for the other fourteen channels The flyback protection devices have one flyback protection fro...

Страница 135: ...he location of the polarization dot as shown in the flyback component locator diagram Figure 5 45 MOV Metal Oxide Varistor or back to back zener diodes are recommended for flyback voltage protection P...

Страница 136: ...20mS 450 volts Capacitance 17 pF Typical Transzorb specifications are Breakdown voltage 300V max at 1mA Reverse Standoff voltage 250 volts Maximum reverse leakage 5mA Maximum peak pulse current 5A Ma...

Страница 137: ...to the J1 connector and loads 9 through 16 connect to the J2 connector Each pin is rated for 2A continuous The following figures show details to help you correctly wire loads to the connectors Figure...

Страница 138: ...h Load Unit User Manual 5 Using Load Cards and Loads CAUTION Connecting high current loads 2 amp per pin can cause premature pin failure Be sure to read Current Sharing of this manual before wiring hi...

Страница 139: ...A C E B D A C E B D 8Pwr J1 Load 8 1 A C E Chan8 D B Load 8 2 A C E B D A C E B D A C E B D 9Pwr Load 9 1 A C E Chan9 D B Load 9 2 A C E B D A C E B D A C E B D 16Pwr Load 16 1 A C E Chan16 D B Load...

Страница 140: ...wr 5Load 1 5Load 2 5Pwr Chan5 5Pwr 6Load 1 6Load 2 6Pwr Chan6 6Pwr 7Load 1 7Load 2 7Pwr Chan7 7Pwr 8Load 1 8Load 2 8Pwr Chan8 8Pwr J2 Connector Wiring Terminal View Row 30 Row 31 Row 32 Row 21 Row 20...

Страница 141: ...1 4Load 1 20 12Load 1 12Load 1 12Load 1 12Load 1 12Load 1 19 4Load 2 4Load 2 4Load 2 4Load 2 4Load 2 19 12Load 2 12Load 2 12Load 2 12Load 2 12Load 2 18 4Pwr 4Pwr 4Pwr 4Pwr 4Pwr 18 12Pwr 12Pwr 12Pwr 12...

Страница 142: ...ns in each row of the connectors see Figure 5 52 This ensures current sharing across all pins and prevents premature pin failure from the excessive current flow External Load Mounting Options As with...

Страница 143: ...is typically connected via cables to a mass interconnect panel User connections to the UUT are then made from the mass interconnect panel Refer to the appropriate mass interconnect wiring guide for c...

Страница 144: ...number on the input is brought back out to the input allowing each channel to operate in a general purpose GP configuration This lets you switch in an external power supply while bypassing the power...

Страница 145: ...Relay Slot Decode Logic P1 PB1 PB2 PB3 PB4 PS 1 PS 2 PS 3 GPIB J3 Configuration Connector External Mount Loads NO1 NC1 Pwr1 Load Switches NO24 Pwr24 NC24 Chan1 Current Sense Control Address Data Power...

Страница 146: ...oad10 4 Load15 1 Load15 2 Load15 3 Load15 4 Load16 1 Load16 2 Load16 3 Load16 4 Load21 3 Load21 4 Load22 3 Load22 4 Load5 1 Load5 2 Load5 3 Load5 4 Load6 1 Load6 2 Load6 3 Load6 4 Load11 3 Load11 4 Lo...

Страница 147: ...er The NO pins are also jumpered to power bus 2 and the NC pins are jumpered to power bus 1 You can easily cut the combs for J802 J807 and J808 J813 to provide bus or pin isolation between the various...

Страница 148: ...he load card You may need to hold the load card at an angle to see the printing from the front when the mezzanine cards are installed The Mezzanine ID pins are located in pairs numbered 1 4 at the top...

Страница 149: ...the card from the connectors especially when multiple cards are installed Use the indentations on the sides of the card to loosen the card with the prying tool then grasp the card and gently rock it...

Страница 150: ...the load card is typically connected via cables to a mass interconnect panel User connections to the UUT are then made from the mass interconnect panel Refer to the appropriate mass interconnect wirin...

Страница 151: ...Using Load Cards and Loads 5 E6198B Switch Load Unit User Manual 5 83 Figure 5 58 Agilent N9378A UUT Connector...

Страница 152: ...mber on the input is brought back out to the input allowing each channel to operate in a general purpose GP configuration This lets you switch in an external power supply while bypassing the power bus...

Страница 153: ...3 GPI B J3 Configuration Connector J801 NC J802 J807 NO2 4 NC24 Mezzanine Card Loads NO1 NC1 Load Switches Pwr1 2 NO 24 2 1 24 2 1 5A fuses User installed jumpers Com1 Chan1 Com2 Chan2 Load1 1 Load1 2...

Страница 154: ...ad22 2 Load22 1 Load21 2 Load21 1 Load19 2 Load19 1 PB1 PB2 PB4 PB3 24 NC 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 0 NO Ch7 Ch8 Mezzanine Card 2 Uppper Mezzanine Card 3 Uppper Mezzani...

Страница 155: ...ected to any of the four power busses on J801 via jumper wires Figure 5 54 The terminal block jumpering allows convenient pull up or pull down of the various inputs It also lets you terminate a UUT lo...

Страница 156: ...r Switches as General Purpose Relays for more information Connecting Loads Loads are mounted on the Agilent N9379A s mezzanine cards Figure 5 62 Up to 16 loads can fit on each mezzanine card with two...

Страница 157: ...anine cards are installed The Mezzanine ID pins are located in pairs numbered 1 4 at the top of the card This lets you identify each mezzanine card using a four digit binary number least significant d...

Страница 158: ...5 90 E6198B Switch Load Unit User Manual 5 Using Load Cards and Loads THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 159: ...igure and use the Agilent Pin Matrix Cards which slots into the Switch Load unit Register descriptions for these cards are located in Appendix B of this manual NOTE This chapter describes using the Ag...

Страница 160: ...imply add more Agilent E8793A 32 Pin Matrix Cards to the bus Features Key features of the cards include 16 x 5 high speed reed relay Instrument Matrix Agilent E8793A only 32 x 4 high speed reed relay...

Страница 161: ...gilent E8793A also contains two additional sets of Analog Bus access on connector J1 Other than these differences the two cards are electrically and functionally identical The following discussion app...

Страница 162: ...s or other user defined applications For example you can close any of these auxiliary relays to connect a digital sensing source or other low impedance system resource such an event detector or digita...

Страница 163: ...OM 96 Pin DIN C To Instruments 96 Pin DIN C Backplane Access To Switch Load Unit 96 Pin DIN C Row Aux UUT Common To UUT 96 Pin DIN C Rear Aux Access Disconnect Relays J2 J1 J3 Bypass Relays 32 x 4 Mea...

Страница 164: ...ector Decoding and Driver Circuits 96 Pin DIN C Backplane Access To Switch Load Unit 96 Pin DIN C Row Aux UUT Common To UUT 96 Pin DIN C Rear Aux Access Disconnect Relays J2 J3 Bypass Relays 32 x 4 Me...

Страница 165: ...he disconnect relays are either under manual or automatic control When in automatic mode when you close a matrix relay the disconnect relay associated with that column also closes When in manual mode...

Страница 166: ...ed to bypass the series protection resistors You can bypass the protection on a column by column basis Reset State The card resets to its default state whenever Operating power is first applied Operat...

Страница 167: ...nnel who are aware of the hazards involved should install remove or configure the Switch Load Unit or plug in cards Before you remove any installed card disconnect AC power from the mainframe and from...

Страница 168: ...nd INST7 INST8 25 Gnd INST8 INST9 24 Gnd INST9 INST10 23 Gnd INST10 INST11 22 Gnd INST11 INST12 21 Gnd INST12 INST13 20 Gnd INST13 INST14 19 Gnd INST14 INST15 18 Gnd INST15 INST16 17 Gnd INST16 N C 16...

Страница 169: ...1 Example Instrument Connections Agilent E3750 61621 E6170 61612 Floating Instruments Agilent E3750 61604 E6170 61614 Single Ended Instruments No Connection Signal Drain Signal Low No Connection INST2...

Страница 170: ...6 12 E6198B Switch Load Unit User Manual 6 Using the Pin Matrix Cards Figure 6 7 shows two typical BNC cables used for instrument connetions to J1 J2 Connector Pinouts Figure 6 7 Instrument BNC Cables...

Страница 171: ...23 UUT Common System Ground Aux23 22 UUT Common System Ground Aux22 21 UUT Common System Ground Aux21 20 UUT Common System Ground Aux20 19 UUT Common System Ground Aux19 18 UUT Common System Ground Au...

Страница 172: ...UUT Common Aux7 Row7 8 UUT Common Aux8 Row8 9 UUT Common Aux9 Row9 10 UUT Common Aux10 Row10 11 UUT Common Aux11 Row11 12 UUT Common Aux12 Row12 13 UUT Common Aux13 Row13 14 UUT Common Aux14 Row14 15...

Страница 173: ...Card can be installed in any available Switch Load Unit slot The Analog Bus connection cables require that all 32 Pin Matrix Cards be in adjacent slots so the cables will reach Figure 6 10 shows a ty...

Страница 174: ...nnections simply add more Agilent E8783A Pin Matrix Cards to the bus Features Key features of the cards include 24 x 5 high speed reed relay Instrument Matrix and 40 x 4 high speed reed relay Measurem...

Страница 175: ...Matrix It also contains two additional sets of Analog Bus access on connector J1 The Agilent E8783A contains 64 x 4 Measurement Matrix only Features Common to Both Cards The Pin Matrix Modules contain...

Страница 176: ...user defined applications For example you can close any of these auxiliary relays to connect a digital sensing source or other low impedance system resource such an event detector or digital input car...

Страница 177: ...ux39 Aux40 160 Pin DIN C Rear Aux Access J2 J1 96 Pin DIN C To Instrument Abus Access Abus Connector J3 Aux1 Aux2 Aux3 Aux38 Aux39 Aux40 Decoding and Driver Circuits DAC1 DAC2 DACCOM P1 96 Pin DIN C B...

Страница 178: ...Aux62 Aux63 Aux64 160 PinDINC Rear AuxAccess J2 AbusConnector J3 Aux1 Aux2 Aux3 Aux62 Aux63 Aux64 DecodingandDriver Circuits P1 96 PinDINC BackplaneAccess ToSwitch LoadUnit P2 160 PinDINC Row Aux UUT...

Страница 179: ...the Column Control Protection Bypass Register When executed OAR re triggers the relay timer Reset Bit 0 in the Control Register programmatically resets the card Resetting the card clears all internal...

Страница 180: ...ollowing pages shows the pinouts for the Agilent E8782A and E8783A user connectors Figure 6 14 Column Disconnect and Bypass Protection Relays WARNING SHOCK HAZARD Only service trained personnel who ar...

Страница 181: ...INST5 INST6 27 Gnd INST6 INST7 26 Gnd INST7 INST8 25 Gnd INST8 INST9 24 Gnd INST9 INST10 23 Gnd INST10 INST11 22 Gnd INST11 INST12 21 Gnd INST12 INST13 20 Gnd INST13 INST14 19 Gnd INST14 INST15 18 Gn...

Страница 182: ...ure 6 6 For E8782A the instrument number extends to instrument 24 The typical BNC cables used for instrument connections to J1 J2 Connector Pinouts are shown in Figure 6 7 Figure 6 16 shows the pinout...

Страница 183: ...m Ground NC NC 21 Aux12 UUT Common System Ground NC NC 20 Aux13 UUT Common System Ground NC NC 19 Aux14 UUT Common System Ground NC NC 18 Aux15 UUT Common System Ground NC NC 17 Aux16 UUT Common Syste...

Страница 184: ...ux14 UUT Common System Ground UUT Common AUX 46 18 Aux15 UUT Common System Ground UUT Common AUX 47 17 Aux16 UUT Common System Ground UUT Common AUX 48 16 Aux17 UUT Common System Ground UUT Common AUX...

Страница 185: ...x9 NC UUT Common 10 Row 10 NC Aux10 NC UUT Common 11 Row 11 NC Aux11 NC UUT Common 12 Row 12 NC Aux12 NC UUT Common 13 Row 13 NC Aux13 NC UUT Common 14 Row 14 NC Aux14 NC UUT Common 15 Row 15 NC Aux15...

Страница 186: ...Aux13 Aux45 UUT Common 14 Row 14 Row 46 Aux14 Aux46 UUT Common 15 Row 15 Row 47 Aux15 Aux47 UUT Common 16 Row 16 Row 48 Aux16 Aux48 UUT Common 17 Row 17 Row 49 Aux17 Aux49 UUT Common 18 Row 18 Row 50...

Страница 187: ...Matrix Card can be installed in any available Switch Load Unit slot The Analog Bus connection cables require that all 64 Pin Matrix Cards be in adjacent slots so the cables will be able to reach each...

Страница 188: ...6 30 E6198B Switch Load Unit User Manual 6 Using the Pin Matrix Cards THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 189: ...r Breakouts 7 4 E8794A Components and Schematics 7 8 The Agilent E8794A Custom Card provides a general purpose breadboard card for system integrators to add custom circuitry The card can also emulate...

Страница 190: ...nnels In addition 32 configuration lines are brought from the custom card to the TC connectors on the Test System Interface Figure 7 1 shows the J2 J6 connectors used to route channels and signals to...

Страница 191: ...neral purpuse breakout input J3 and J4 route DAC and or Event Detector lines 96 Pin DIN C connector Routes Config Lines to Test System Interface 96 Pin DIN C connector to backplane Brings 12 V Data Ad...

Страница 192: ...1604 is used to connect J2 to the Test System Interface WARNING STATIC ELECTRICITY Static electricity is a major cause of component failure To prevent damage to electrical components observe anti stat...

Страница 193: ...3 1 CH12 LS CH12 LO CH12 HI CH12 HS CH11 LS CH11 LO CH11 HI CH11 HS CH10 LS CH10 LO CH10 HI CH10 HS CH9 LS CH9 LO CH9 HI CH9 HS CH4 LS CH4 LO CH4 HI CH4 HS CH3 LS CH3 LO CH3 HI CH3 HS CH2 LS CH2 LO CH...

Страница 194: ...left breakout connections or J4 right breakout connections Pin Number Signal Pin Number Signal 64 62 60 58 56 54 52 50 48 46 44 42 40 38 36 34 32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2 63 61 59 57 5...

Страница 195: ...tor breakouts when cable part number E3751 61601 is used to connect the 32 Pin Matrix Card s to J5 and J6 Figure 7 5 J5 and J6 Breakouts 32 Pin Matrix Cards Aux32 J6 UUTCommon J5 J6 J5 Note Pinout for...

Страница 196: ...omponents and Schematics Figure 7 6 shows the E8794 layout Figure 7 7 shows a more detailed view for locating components Figure 7 9 shows a schematic view of the E8794A custom card Figure 7 9 shows a...

Страница 197: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 9 Figure 7 7 E8794A Component Locator...

Страница 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...

Страница 199: ...Using the Custom Card 7 E6198B Switch Load Unit User Manual 7 11 Figure 7 9 Status and Control Interface...

Страница 200: ...7 12 E6198B Switch Load Unit User Manual 7 Using the Custom Card THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 201: ...4 Agilent E6177A 24 Channel Medium Current Load Card Specifications A 5 Agilent U7177A 24 Channel Medium Current Load Card Specifications A 6 Agilent E6178B 8 Channel Heavy Duty Load Card Specificati...

Страница 202: ...imum Combined Power Bus Current all buses 60 A maximum continuous Switch Load Unit to Power Supply Cable Resistance per bus 0 03 standard supplied Agilent cable with connectors included Switch Load Un...

Страница 203: ...0 msec maximum 30 cps maximum Peak voltage to Earth 60 V continuous 500 V maximum transient non switching delays to 60V in 200 mS Load carry switching current 7 5 A maximum non switching continuous Lo...

Страница 204: ...operate release time 10 8 msec typical 16 10 msec maximum 30 cps max Peak voltage to Earth 60 V continuous 500 V maximum transient non switching delays to 60V in 200 mS Load carry switching current 7...

Страница 205: ...of load with standard 0 05 sense resistor and 10 A slo blo fuse installed 5 for load currents 0 1 A Load path relay operate release time 3 3 2 4 msec typical 8 8 msec max 10 cps maximum Peak voltage t...

Страница 206: ...elease time 3 3 2 4 msec typical 8 8 msec max 10 cps maximum Peak Voltage to Earth 60 V continuous 500 V maximum transient non switching delays to 60V in 200 mS Load carry switching current 2 A maximu...

Страница 207: ...ed 120 A maximum 10 mS 14 Vdc resistive Power Bus Carry Current continuous 30 A maximum Load Switching Power 500 W max 25Vdc max resistive load 100 W max 40Vdc max resistive load Load Switching Voltag...

Страница 208: ...mum transient non switching decays to 60V in 200mS Load carry switching current 7 5 A maximum non switching continuous Load carry peak current 15 A max non switching 100 msec 2 duty cycle Load switchi...

Страница 209: ...d connect 0 5 maximum Exclusive of load Load path relay operate release time 4 4 msec max 10 cps maximum Peak voltage to Earth 60 V continuous 500 V maximum transient non switching decays to 60V in 20...

Страница 210: ...connect 0 5 maximum Exclusive of load Load path relay operate release time 4 4 msec max 10 cps maximum Peak voltage to Earth 60 V continuous 200 V maximum transient non switching decays to 60V in 200...

Страница 211: ...ed Bandwidth 10 MHz Minimum Balanced Pair Bandwidth 5 MHz Minimum Parameter Specification Power Requirements Voltage 5 Vdc Capacitance DUT pin to UUT Common Open channel 100 pF Closed channel 300 pF R...

Страница 212: ...racteristics Agilent E8792A and E8793A Parameter Specification Type Dry reed Switching Speed Close 500 ms Open 400 ms Switching Characteristics 1 0 A carry 0 5 A while switching 7 5 Volt Amps max inst...

Страница 213: ...ance 1 Unbalanced Pair Bandwidth 10MHz Minimum Balanced Pair Bandwidth 5MHz Minimum Parameter Specification Power Requirements Voltage 5 Vdc Capacitance DUT pin to UUT Common Open channel 100 pF Close...

Страница 214: ...reduce problems caused by relay failure Use the methods below to determine when a relay may be near failure The best method or combination of methods and the failure criteria depend on the application...

Страница 215: ...of life no load Replacement Strategy The best strategy depends on your application Replace individual relays if they are used more often or at higher loads than others If all the relays see similar l...

Страница 216: ...A 16 E6198B Switch Load Unit User Manual A Switch Load Unit And Card Specifications THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 217: ...83 This appendix provides register based programming information for the Agilent E6198B Switch Load Unit backplane the individual load cards the pin cards and the custom card This information can hel...

Страница 218: ...ree registers as show in Table B 1 Table B 2 shows the values returned from each card s Card Type register Table B 1 Standard Registers Register Offset Register Name Description 00h Card type See Tabl...

Страница 219: ...n slot 20 of the first Switch Load Unit would be Frame 1 0012 Slot 20 101002 Register 0 000000002 Base Address 00110100000000002 or 3400h Agilent E8792A 1010 0Ah Fully loaded pin card includes instrum...

Страница 220: ...000002 on the Switch Load Unit selected Table B 3 summarizes the Switch Load Unit registers Table B 3 Switch Load Unit Registers Register Offset Definitions Type Base 0h Card Type Read Only Base 1h Ca...

Страница 221: ...tus register provides readback of the current Reset and Busy status of the backplane Busy is an open collector line that any slot can drive to indicate its status typically a relay timer The state of...

Страница 222: ...Spare_DigIn 0 Spare_DigIn 7 of the system resource access connector J104 DAC1 Output MSB W Base 8h Writing to the DAC1 Output register sets the Most Significant Bit MSB of the DAC1 digital input DAC1...

Страница 223: ...1 to this register wait 5 mS and then write a 0 to this register DACs will reset to 0 Vout NOTE To set the value of the DAC output always write the MSB first followed by the LSB register offset 9h Th...

Страница 224: ...e_DigOut 7 outputs will directly reflect the contents of this register 1 high 0 low DAC2 Output MSB W Base Dh Writing to the DAC2 Output register sets the Most Significant Bit MSB of the DAC2 digital...

Страница 225: ...ut register sets the Least Significant Bit LSB of the DAC2 digital input NOTE To set the value of the DAC output always write the MSB first followed by the LSB The output of the DAC will not update un...

Страница 226: ...control Status control Base 3h Current select Current select Load select 1 8 Current Sense Current select Current select Mezzanin e card 1 config Mezzanin e card 1 config Base 4h Load select 1 8 Load...

Страница 227: ...the card follow Base Ch N A N A N A N A N A N A Load select Load select Base Dh N A N A N A N A N A N A Load select Load select Base Eh N A N A N A N A N A N A Load select Load select Base Fh N A N A...

Страница 228: ...er reads back the card s status Currently the status is defined as the state of the relay timer The relay timer has two timers wire OR d together One timer is designed for the slower armature relays 1...

Страница 229: ...1112 unselected 0012 corresponds to current sense select on channels 1 and 2 0102 corresponds to current sense select on channels 3 and 4 0112 corresponds to current sense select on channels 5 and 6...

Страница 230: ...rm C double pole single throw and the register uses positive logic 1 Normally open NO shorted to COM normally closed NC is open 0 NC shorted to COM NO is open The power buses selected depend on how th...

Страница 231: ...ack the Card Type 02h of the card Read Only Power On Reset State 02h Card Configuration R Base 1h This register reads back the Card Configuration The Card Configuration is determined by the user to di...

Страница 232: ...are not used Valid current sense selection values are 000012 100002 Selecting values 000002 or 100012 111112 will NOT select current sensing for any channel 000012 corresponds to current sense select...

Страница 233: ...armature relay timer Write Only State 1 closed 0 open Power On Reset State 0 Power Select 1 8 W Base 6h This register controls the Pull Up Down Power Select armature relays for channels 1 8 These rel...

Страница 234: ...orm C and the register uses positive logic 1 NO shorted to COM NC is open 0 NC shorted to COM NO is open The power buses selected depend upon how the card is configured by the user Writing to this reg...

Страница 235: ...the card follow Card Type R Base 0h This register reads back the Card Type 03h of the card Read Only Power On Reset State 03h Card Configuration R Base 1h This register reads back the Card Configurati...

Страница 236: ...r starts the relay timer Write Only State 1 closed 0 open Power On Reset State 0 Load Select 9 16 W Base 4h This register controls the Load Select switch armature relays for channels 9 16 The register...

Страница 237: ...nnels 1 8 These relays are Form C and the register uses positive logic 1 NO shorted to COM NC is open 0 NC shorted to COM NO is open The power buses selected are determined by how the card is configur...

Страница 238: ...s register controls the Pull Up Down Power Select armature relays for channels 17 24 These relays are Form C and the register uses positive logic 1 NO shorted to COM NC is open 0 NC shorted to COM NO...

Страница 239: ...the Card Type 18h of the card Read Only Power On Reset State 18h Card Configuration R Base 1h This register reads back the Card Configuration The Card Configuration is determined by the user to distin...

Страница 240: ...current sense relay to be selected Writing to this register starts the reed relay timer Write Only Undefined bits are not used Valid current sense selection values are 0000012 0110002 Selecting values...

Страница 241: ...eset State 0 Load Select 9 16 W Base 5h This register controls the Load Select switch armature relays for channels 9 16 The register uses positive logic 1 closed Writing to this register starts the re...

Страница 242: ...ese relays are Form C and the register uses positive logic 1 NO shorted to COM NC is open 0 NC shorted to COM NO is open The power buses selected are determined by how the card is configured Writing t...

Страница 243: ...orted to COM NC is open 0 NC shorted to COM NO is open Power On Reset State 0 Power Select 17 24 W Base 9h This register controls the Pull Up Down Power Select armature relays for channels 17 24 These...

Страница 244: ...inuous current on one channel at a time See Using the Agilent U7177A 24 Channel Load Card for more information Card Type R Base 00h This register reads back the Card Type 04h of the card Card Configur...

Страница 245: ...selects are encoded This allows for a single channel to be selected At reset no channel is selected Load Select W Base 04h The register controls the Load Select switch armature relays K1 K8 of the car...

Страница 246: ...relay and writing a 0 opens de energizes the relay The reset and power on state for all relays is open The block diagram in Figure shows the location of the various registers CAUTION It is possible t...

Страница 247: ...D to read this register Note also that the N9377A load card supports a CC8 bit for offset register base 2h See Status Card Configuration Bit 8 R Base 2h Read Only Undefined bits read back as all 1s FF...

Страница 248: ...per slot is allowed these channel selects are encoded This allows selection of zero or only one current sense relay at a time Writing to this register starts the reed relay timer Write Only Undefined...

Страница 249: ...adx to Chanx for channels 9 16 See Figure B 10 Write Only State 0 open 1 closed Power On Reset State 0 all channels open Power Select 1 8 W Base 6h This register controls the Pull Up Down Power Select...

Страница 250: ...power buses selected depend upon how the card is configured by the user Writing to this register starts the armature relay timer Write Only State 1 NO shorted to Pwr node State 0 NC shorted to Pwr no...

Страница 251: ...ent N9378A low resistance load card lets you attach as many as four loads per channel Loads are mounted onboard on a series of mezzanine cards that plug into the load card The N9378A card also provide...

Страница 252: ...tInfo to read this register Read Only x undefined Status Control R W Base 2h This register reads back the card s status It also can be used to reset the board or open all relays OAR Card status is def...

Страница 253: ...ed Mezzanine Card 1 Configuration R Base 3h This register reads configuration information from the 1 upper and 1 lower mezzanine cards Card configuration is determined by the user to distinguish diffe...

Страница 254: ...its 4 7 1 lower card MC1L 0 MC1L 3 Power On Reset State 1 Mezzanine Card 2 Configuration R Base 4h This register reads configuration information from the 2 upper and 2 lower mezzanine cards See Figure...

Страница 255: ...information from the 3 upper and 3 lower mezzanine cards Card configuration is determined by the user to distinguish different load configurations There are four bits available for each card which pe...

Страница 256: ...osed 0 open Power On Reset State 0 Load Select 5 6 W Base 8h This register controls the power selection relays for channels 5 and 6 with four possible loads on each channel Bit 7 6 5 4 3 2 1 0 Purpose...

Страница 257: ...State 0 Load Select 9 10 W Base Ah This register controls the power selection relays for channels 10 and 9 with four possible loads on each channel Write Only State 1 closed 0 open Power On Reset Sta...

Страница 258: ...n Power On Reset State 0 Load Select 15 16 W Base Dh This register controls the power selection relays for channels 16 and 15 with four possible loads on each channel Bit 7 6 5 4 3 2 1 0 Purpose Chan1...

Страница 259: ...e 0 Load Select 19 20 W Base Fh This register controls the power selection relays for channels 20 and 19 with four possible loads on each channel Write Only State 1 closed 0 open Power On Reset State...

Страница 260: ...each channel Write Only State 1 closed 0 open Power On Reset State 0 Power Select 1 8 W Base 12h This register controls the Pull Up Down Power Select armature relays for channels 1 8 These relays are...

Страница 261: ...0 NC shorted to COM NO is open The power buses selected are determined by how the card is configured Writing to this register starts the relay timer Write Only State 1 NO shorted to Pwr node State 0...

Страница 262: ...ation Loads are mounted onboard on a series of mezzanine cards that plug into the load card You can attach as many as two loads per channel See Using the Agilent N9379A 48 Channel High Density Load Ca...

Страница 263: ...e card s status It also can be used to reset the board or open all relays OAR Card status is defined as the state of the relay timer The relay timer is designed for the armature relays 4ms The timer r...

Страница 264: ...s available for each card which permits identification of the mezzanine card location and proper orientation Use the action loadCardGetInfo to read this register Figure B 12 shows the layout of the me...

Страница 265: ...fault read back is FFh but you can pull down bits to zero as required to create a configuration byte unique to the loads installed Read Only Lower nibble bits 0 3 1 upper card MC2U 0 MC2U 3 Upper nibb...

Страница 266: ...ach channel Write Only State 1 closed 0 open Power On Reset State 0 Load Select 5 8 W Base 7h This register controls the power selection relays for channels 5 through 8 with two possible loads on each...

Страница 267: ...Select 13 16 W Base 9h This register controls the power selection relays for channels 13 through 16 with two possible loads on each channel Write Only State 1 closed 0 open Power On Reset State 0 Bit...

Страница 268: ...pen Power On Reset State 0 Load Select 25 28 W Base Ch This register controls the power selection relays for channels 25 through 28 with two possible loads on each channel Bit 7 6 5 4 3 2 1 0 Purpose...

Страница 269: ...e 0 Load Select 33 36 W Base Eh This register controls the power selection relays for channels 33 through 36 with two possible loads on each channel Write Only State 1 closed 0 open Power On Reset Sta...

Страница 270: ...pen Power On Reset State 0 Load Select 44 48 W Base 11h This register controls the power selection relays for channels 45 through 48 with two possible loads on each channel Bit 7 6 5 4 3 2 1 0 Purpose...

Страница 271: ...to Pwr node Power On Reset State 0 all channels connected to NC Power Select 17 32 W Base 13h This register controls the Pull Up Down Power Select armature relays for channels 17 32 Each bit controls...

Страница 272: ...o channels These relays are Form C and the register uses positive logic 1 NO shorted to COM NC is open 0 NC shorted to COM NO is open The power buses selected are determined by how the card is configu...

Страница 273: ...92A returns a card type of 1010 dddd 1010 Agilent E8793A returns card type of 1110 dddd 1011 Card Configuration R Base 1h Pin card configuration register and firmware revision FWR register Initial fir...

Страница 274: ...ed to low side of DMM ISensRly 0 opens relay from Inst16 Spare16 to ISense default 1 closes relay from Inst16 Spare16 to Isense OAR 0 causes no change 1 causes all relays to open NOTE See description...

Страница 275: ...elays Not Used Base 5h NOTE The only difference between OAR and Reset is OAR leaves the state of Manual the same Reset clears the Manual setting NOTE Register offset 3h used only for a placeholder to...

Страница 276: ...9 16 W Base Bh 1110 Bits 7 6 5 4 3 2 1 0 Write Aux8 Aux7 Aux6 Aux5 Aux4 Aux3 Aux2 Aux1 Bits 7 6 5 4 3 2 1 0 Write Aux16 Aux15 Aux14 Aux13 Aux12 Aux1 Aux10 Aux9 Bits 7 6 5 4 3 2 1 0 Write Aux24 Aux23...

Страница 277: ...e 11h 1710 Bits 7 6 5 4 3 2 1 0 Write Row24 Row23 Row22 Row21 Row20 Row19 Row18 Row17 Bits 7 6 5 4 3 2 1 0 Write Row32 Row31 Row30 Row29 Row28 Row27 Row26 Row25 Bits 7 6 5 4 3 2 1 0 Write Row8 Row7 Ro...

Страница 278: ...9 16 W Base 17h 2310 Bits 7 6 5 4 3 2 1 0 Write Row8 Row7 Row6 Row5 Row4 Row3 Row2 Row1 Bits 7 6 5 4 3 2 1 0 Write Row16 Row15 Row14 Row13 Row12 Row11 Row10 Row9 Bits 7 6 5 4 3 2 1 0 Write Row24 Row2...

Страница 279: ...lay 9 16 W Base 1Bh 2710 Abus1 to Instrument Bus Instrument Relay 1 8 W Base 1Ch 2810 Bits 7 6 5 4 3 2 1 0 Write Row24 Row23 Row22 Row21 Row20 Row19 Row18 Row17 Bits 7 6 5 4 3 2 1 0 Write Row32 Row31...

Страница 280: ...ument Bus Instrument Relay 1 8 W Base 22h 3410 Bits 7 6 5 4 3 2 1 0 Write Inst16 Inst15 Inst14 Inst13 Inst12 Inst1 Inst10 Inst9 Bits 7 6 5 4 3 2 1 0 Write Inst8 Inst7 Inst6 Inst5 Inst4 Inst3 Inst2 Ins...

Страница 281: ...atrix card with auxiliary returns a card type of 6710 ddddddd 1000011 Card Configuration R Base 1h Instrument matrix card configuration and firmware revision FWR register Initial firmware revision 0h...

Страница 282: ...Not Used Base 3h Register offset 3h used only for placeholder to make software addressing simpler Abus control and protection bypass relay W Base 4h Instrument matrix card Abus and protection bypass r...

Страница 283: ...ay 9 16 W Base Ah 1010 Auxilary Relay 17 24 W Base Bh 1110 Bits 7 6 5 4 3 2 1 0 Write Inst16 Inst15 Inst14 Inst13 Inst12 Inst11 Inst10 Inst9 Bits 7 6 5 4 3 2 1 0 Write Inst24 Inst23 Inst22 Inst21 Inst...

Страница 284: ...0 Instrument Relay 17 24 W Base 10h 1610 Bits 7 6 5 4 3 2 1 0 Write Aux32 Aux31 Aux30 Aux29 Aux28 Aux27 Aux26 Aux25 Bits 7 6 5 4 3 2 1 0 Write Aux40 Aux39 Aux38 Aux37 Aux36 Aux35 Aux34 Aux33 Bits 7 6...

Страница 285: ...0 Row Relay 33 40 W Base 15h 2110 Bits 7 6 5 4 3 2 1 0 Write Row8 Row7 Row6 Row5 Row4 Row3 Row2 Row1 Bits 7 6 5 4 3 2 1 0 Write Row16 Row15 Row14 Row13 Row12 Row11 Row10 Row9 Bits 7 6 5 4 3 2 1 0 Writ...

Страница 286: ...W Base 19h 2510 Row Relay 9 16 W Base 1Ah 2610 Bits 7 6 5 4 3 2 1 0 Write Inst8 Inst7 Inst6 Inst5 Inst4 Inst3 Inst2 Inst1 Bits 7 6 5 4 3 2 1 0 Write Inst16 Inst15 Inst14 Inst13 Inst12 Inst11 Inst10 I...

Страница 287: ...Instrument Relay 9 16 W Base 1Fh 3110 Bits 7 6 5 4 3 2 1 0 Write Row24 Row23 Row22 Row21 Row20 Row19 Row18 Row17 Bits 7 6 5 4 3 2 1 0 Write Row32 Row31 Row30 Row29 Row28 Row27 Row26 Row25 Bits 7 6 5...

Страница 288: ...se 25h 3710 Bits 7 6 5 4 3 2 1 0 Write Inst24 Inst23 Inst22 Inst21 Inst20 Inst19 Inst18 Inst17 Bits 7 6 5 4 3 2 1 0 Write Row8 Row7 Row6 Row5 Row4 Row3 Row2 Row1 Bits 7 6 5 4 3 2 1 0 Write Row16 Row15...

Страница 289: ...W Base 29h 4110 Row Relay 9 16 W Base 2Ah 4210 Bits 7 6 5 4 3 2 1 0 Write Inst8 Inst7 Inst6 Inst5 Inst4 Inst3 Inst2 Inst1 Bits 7 6 5 4 3 2 1 0 Write Inst16 Inst15 Inst14 Inst13 Inst12 Inst11 Inst10 I...

Страница 290: ...tracking must be done in software Card Type R Base 0h Returns measurement matrix card type Measurement matrix card without auxiliary returns a card type of 7010 ddddddd 1000110 Measurement matrix car...

Страница 291: ...ses board reset and triggers the relay timer Clears itself afterward Not Used Base 3h Register offset 3h used only for placeholder to make software addressing simpler Abus control and protection bypas...

Страница 292: ...16 W Base 7h Auxilary Relay 17 24 W Base 8h Auxilary Relay 25 32 W Base 9h Auxilary Relay 33 40 W Base Ah 1010 Bits 7 6 5 4 3 2 1 0 Write Aux8 Aux7 Aux6 Aux5 Aux4 Aux3 Aux2 Aux1 Bits 7 6 5 4 3 2 1 0...

Страница 293: ...Eh 1410 Row Relay 9 16 W Base Fh 1510 Bits 7 6 5 4 3 2 1 0 Write Aux48 Aux47 Aux46 Aux45 Aux44 Aux43 Aux42 Aux41 Bits 7 6 5 4 3 2 1 0 Write Aux56 Aux55 Aux54 Aux53 Aux52 Aux51 Aux50 Aux49 Bits 7 6 5...

Страница 294: ...10 Bits 7 6 5 4 3 2 1 0 Write Row24 Row23 Row22 Row21 Row20 Row19 Row18 Row17 Bits 7 6 5 4 3 2 1 0 Write Row32 Row31 Row30 Row29 Row28 Row27 Row26 Row25 Bits 7 6 5 4 3 2 1 0 Write Row40 Row39 Row38 Ro...

Страница 295: ...ase 1Bh 2710 Bits 7 6 5 4 3 2 1 0 Write Row8 Row7 Row6 Row5 Row4 Row3 Row2 Row1 Bits 7 6 5 4 3 2 1 0 Write Row16 Row15 Row14 Row13 Row12 Row11 Row10 Row9 Bits 7 6 5 4 3 2 1 0 Write Row24 Row23 Row22 R...

Страница 296: ...ase 21h 3310 Bits 7 6 5 4 3 2 1 0 Write Row56 Row55 Row54 Row53 Row52 Row51 Row50 Row49 Bits 7 6 5 4 3 2 1 0 Write Row64 Row63 Row62 Row61 Row60 Row59 Row58 Row57 Bits 7 6 5 4 3 2 1 0 Write Row8 Row7...

Страница 297: ...ase 27h 3910 Bits 7 6 5 4 3 2 1 0 Write Row40 Row39 Row38 Row37 Row36 Row35 Row34 Row33 Bits 7 6 5 4 3 2 1 0 Write Row48 Row47 Row46 Row45 Row44 Row43 Row42 Row41 Bits 7 6 5 4 3 2 1 0 Write Row56 Row5...

Страница 298: ...10 Bits 7 6 5 4 3 2 1 0 Write Row24 Row23 Row22 Row21 Row20 Row19 Row18 Row17 Bits 7 6 5 4 3 2 1 0 Write Row32 Row31 Row30 Row29 Row28 Row27 Row26 Row25 Bits 7 6 5 4 3 2 1 0 Write Row40 Row39 Row38 Ro...

Страница 299: ...alue configuration Control register W Base 2h Writing a 1 to bit seven generates a reset pulse duration 250 nS which you can use to reset on board circuitry The reset pulse is self clearing Digital In...

Страница 300: ...drain with a light pull up 100 k ohm Writing a 1 sets the output of a open collector driver on meaning that the output voltage is pulled low The reset state of this register is 0h all outputs off This...

Страница 301: ...C 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies C Repair Information Support Strategy C 2 Locating Load Card Components C 2...

Страница 302: ...can determine which relay is defective you can repair the module The relays are through hole technology and easy to replace Locating Load Card Components Agilent E6175A 8 Channel High Current Card Se...

Страница 303: ...Repair Information C E6198B Switch Load Unit User Manual C 3 Figure C 1 Agilent E6175A Component Locator Rear View...

Страница 304: ...the E6176A The component locator diagram Blue Figure shows the solder side or back of the Agilent E3750 66504 through hole PC board This load card uses the 0490 1517 Form C relay Two Agilent 0490 183...

Страница 305: ...n relays Individual relays can be replaced on the Agilent E8792A and E8793A 32 pin matrix modules All relays are 5V 1A reed relays The Agilent part number for an 8 pack of relays is 0490 1838 The tabl...

Страница 306: ...8 K418 K518 Row19 K119 K219 K319 K419 K519 Row20 K120 K220 K320 K420 K520 Row21 K121 K221 K321 K421 K521 Row22 K122 K222 K322 K422 K522 Row23 K123 K223 K323 K423 K523 Row24 K124 K224 K324 K424 K524 Ro...

Страница 307: ...07 K1407 K1007 Inst8 K1108 K1208 K1308 K1408 K1008 Inst9 K1109 K1209 K1309 K1409 K1009 Inst10 K1110 K1210 K1310 K1410 K1010 Inst11 K1111 K1211 K1311 K1411 K1011 Inst12 K1112 K1212 K1312 K1412 K1012 In...

Страница 308: ...lumn3 Abus3 is K305 Relay component numbers are silk screened on the back of the PC board Note that E8782A with ID 6610 and E8783A with ID 7010 do not have Row to Aux relays Table C 5 Measurement Matr...

Страница 309: ...2 K232 K332 K432 K932 Row9 K133 K233 K333 K433 K933 Row10 K134 K234 K334 K434 K934 Row11 K135 K235 K335 K435 K935 Row12 K136 K236 K336 K436 K936 Row13 K137 K237 K337 K437 K937 Row14 K138 K238 K338 K43...

Страница 310: ...3 K363 K463 K963 Row40 K164 K264 K364 K464 K964 Table C 6 Measurement Matrix Relays for E8783A Column1 Abus1 Column2 Abus2 Column3 Abus3 Column4 Abus4 Row to Aux Row1 K101 K201 K301 K401 K901 Row2 K10...

Страница 311: ...27 K227 K327 K427 K927 Row28 K128 K228 K328 K428 K928 Row29 K129 K229 K329 K429 K929 Row30 K130 K230 K330 K430 K930 Row31 K131 K231 K331 K431 K931 Row32 K132 K232 K332 K432 K932 Row33 K133 K233 K333 K...

Страница 312: ...ow55 K155 K255 K355 K455 K955 Row56 K156 K256 K356 K456 K956 Row57 K157 K257 K357 K457 K957 Row58 K158 K258 K358 K458 K958 Row59 K159 K259 K359 K459 K959 Row60 K160 K260 K360 K460 K960 Row61 K161 K261...

Страница 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...

Страница 314: ...back up to the UUT load side of the card and jumpering a channel s internal load terminals allows the user to utilize the load switch as a GP relay Current Sensing Determining the current through a f...

Страница 315: ...removing jumper plugs on JP1 JP2 and JP3 Current sensing is performed on any load card channel across a four terminal current sense resistor Two load cards are designed to connect to the current sense...

Страница 316: ...tacts A Form C Single Pole Double Throw switch has two possible states The default or un powered state is its normal state The two terminals on the switch are therefore called normally open or normall...

Страница 317: ...ment or system on itself to verify the functionality of the instrument or system Sense See Current sensing Slot Decode Logic The circuitry that interprets the data that represents the physical slot a...

Страница 318: ...G 6 E6198B Switch Load Unit User Manual G Glossary Of Terms THIS PAGE IS INTENTIONALLY LEFT BLANK...

Отзывы: