Appendix Service Information
Constant Voltage (CV) Verifications
172
1
Turn off the power supply and connect the output to be tested as shown in
Figure A-1 to an oscilloscope (ac coupled) between (+) and (-) terminals. Set
the oscilloscope to AC mode and bandwidth limit to 20 MHz. Connect a
resistive load (13.5
9
)
‡
to the terminal at (B) as shown above.
2
Turn on the power supply. Select the high voltage range (20V/1.5A)
†
, enable
the output, and set the display to the limit mode. When the display is in the
limit mode, program the current to the full scale rated value (1.5 A)
†
and the
voltage to the full rated value (20.0 V)
†
.
3
Check that the front panel
CV
annunciator remains lit. If not lit, adjust the load
down slightly.
4
Note that the waveform on the oscilloscope does not exceed the peak-to-peak
limit of (5 mV)* or (8 mV)**.
5
Disc,onnect the oscilloscope and connect the ac rms voltmeter in its place
according to the VRMS measurement techniques above and as shown above.
The rms voltage reading does not exceed the rms limit of 0.5 mV* or 1 mV**.
N o t e :
For better measurement result, it is recommended to make the connection between
the BNC receptacle and the output terminals shorter as much as possible, and to use
the recommended split ferrites with the cable (BNC to BNC) as shown above.
Load Resistor
Input
RMS voltmeter
Load Resistor
Input
RMS voltmeter
(Front Panel Connections)
(Rear Panel Connections)
BNC Cable
BNC
Receptacle
Split Ferrites*
*For E3640A/42A/44A models. **For E3641A/43A/45A models.
†
For E3640A model, and see
Table A-3
for other models.
‡
For E3640A model, and see
Table A-2
for other models.
Содержание E364 A Series
Страница 9: ...8...
Страница 15: ...14 Contents Contents...
Страница 16: ...1 Quick Start...
Страница 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Страница 28: ...2 General Information...
Страница 40: ...3 Front Panel Operation and Features...
Страница 68: ...4 Remote Interface Reference...
Страница 114: ...5 Error Messages...
Страница 124: ...6 Application Programs...
Страница 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Страница 136: ...7 Tutorial...
Страница 147: ...Chapter 7 Tutorial Remote Programming 146...
Страница 148: ...8 Specifications...
Страница 155: ...Chapter 8 Specifications Supplemental Characteristics 154 Figure 8 2 Dimensions for Rack mounting...
Страница 156: ...Appendix Service Information...
Страница 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Страница 193: ...Appendix Service Information Replaceable Parts 192...
Страница 197: ......
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......