Appendix Service Information
Constant Current (CC) Verifications
175
Ser
vice
In
for
mat
ion
6
Record the value displayed on the controller. This value should be within the
limit of (I
O
± 5 mA).
7
Program the output current to the full scale rated value (3 A)
†
by sending the
command:
CURR 3.0
(E3640A model)
8
Divide the voltage drop (DVM reading) across the current monitoring resistor
(R
M
) by its resistance to convert to amps and record this value (I
O
). This value
should be within the limit of:
9
Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
10
Record the value displayed on the controller. This value should be within the
limit of:
CC Load Effect (Load Regulation)
This test measures the change in output current resulting from a change in the
load from full rated output voltage to short circuit.
1
Turn off the power supply and connect the output to tested as shown in
Figure A-1 with the digital voltmeter connected across the 0.01
9
current
monitoring resistor (R
M1
).
2
Turn on the power supply. Select the low voltage range (8V/3A)
†
, enable the
output, and set the display to the limit mode. When the display is in the limit
mode, program the output voltage to the full scale rated value (8.0 V)
†
and the
output current to the full rated value (3 A)
†
.
3
Set the voltage of the electronic load to (8.0 V)
†
to operate it in constant voltage
mode since a voltage drop occurs on the load wires. Check that the
CC
annunciator is on. If it is not, adjust the load so that the output voltage drops
slightly. Record the current reading by dividing the voltage reading on the
digital voltmeter by the resistance of the current monitoring resistor.
E3640A
E3641A
E3642A
E3643A
E3644A
E3644A
3 A ± 16 mA 0.8 A ± 11.6 mA
5 A ± 20 mA
5 A ± 20 mA
8 A ± 26 mA
8 A ± 26 mA
E3640A
E3641A
E3642A
E3643A
E3644A
E3642A
(I
O
± 9.5 mA)
(I
O
± 6.2 mA) (I
O
± 12.5 mA) (I
O
± 7.1 mA)
(I
O
± 17 mA)
(I
O
± 8.3 mA)
†
For E3640A model, and see
Table A-3
for other models
Содержание E364 A Series
Страница 9: ...8...
Страница 15: ...14 Contents Contents...
Страница 16: ...1 Quick Start...
Страница 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Страница 28: ...2 General Information...
Страница 40: ...3 Front Panel Operation and Features...
Страница 68: ...4 Remote Interface Reference...
Страница 114: ...5 Error Messages...
Страница 124: ...6 Application Programs...
Страница 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Страница 136: ...7 Tutorial...
Страница 147: ...Chapter 7 Tutorial Remote Programming 146...
Страница 148: ...8 Specifications...
Страница 155: ...Chapter 8 Specifications Supplemental Characteristics 154 Figure 8 2 Dimensions for Rack mounting...
Страница 156: ...Appendix Service Information...
Страница 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Страница 193: ...Appendix Service Information Replaceable Parts 192...
Страница 197: ......
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......