Appendix Service Information
Self-Test Procedures
163
Ser
vice
In
for
mat
ion
606
Rundown gain out of range
This test checks the nominal gain between the
integrating ADC and the U121 on-chip ADC. The nominal gain is checked to
±10% tolerance.
607
Rundown too noisy
This test checks the gain repeatability between the
integrating ADC and the U121 on-chip ADC. The gain test (606) is performed
eight times. Gain noise must be less that ±64 lsb’s of the U121 on-chip ADC.
608
Serial configuration readback failed
This test re-sends the last 3 byte serial
configuration data to all the serial path (SERDAT, SERBCK, SERCLK). The
data is then clocked back into U130 and compared against the original 3 bytes
sent. A failure occurs if the data do not match. This tests checks the serial data
path through U138.
624
Unable to sense line frequency
This test checks that the LSENCE logic
input U121 is toggling. If no logic input detected, the power supply will assume
a 50 Hz line operation for all future measurements.
625
I/O processor did not respond
This test checks that communications can
be established between U121 and U103 through the optically isolated (U108
and U109) serial data link. Failure to establish communication in either
direction will generate an error. If this condition is detected at power-on
self-test, the power supply will beep and the error annunciator will be on.
626
I/O processor failed self-test
This test causes the earth referenced
processor U103 to execute an internal, ram test. Failure will generate an error.
630
Fan test failed
This test checks if the fan current is flowing. If the current
is not detected at power-on self-test, the power supply will beep and the error
annunciator will be on. Fan test fail could likely induce overtemperature
condition in the power supply.
631
System DAC test failed
This test checks if the DAC hardware is functional.
The main controller U121 sends a reference voltage data to DAC and converts
the DAC output to digital data to see if the digital data is within a valid range.
632
Hardware test failed
This test checks the status of voltage and current error
amplifiers for the power circuit. If both amplifiers are not operational, the
power supply will beep and the error annunciator will be on.
Содержание E364 A Series
Страница 9: ...8...
Страница 15: ...14 Contents Contents...
Страница 16: ...1 Quick Start...
Страница 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Страница 28: ...2 General Information...
Страница 40: ...3 Front Panel Operation and Features...
Страница 68: ...4 Remote Interface Reference...
Страница 114: ...5 Error Messages...
Страница 124: ...6 Application Programs...
Страница 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Страница 136: ...7 Tutorial...
Страница 147: ...Chapter 7 Tutorial Remote Programming 146...
Страница 148: ...8 Specifications...
Страница 155: ...Chapter 8 Specifications Supplemental Characteristics 154 Figure 8 2 Dimensions for Rack mounting...
Страница 156: ...Appendix Service Information...
Страница 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Страница 193: ...Appendix Service Information Replaceable Parts 192...
Страница 197: ......
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......