Instrument Installation Checklist
Revision:
A.00
, Issued:
January
17
, 20
21
Document part number:
D0007550
Page ____ of ____
© Agilent Technologies, Inc. 20
21
g
Close the Robot Teaching dialog box.
2
Remove the stage jig from the robot grippers and turn off the power on the jig.
3
Reset the config file so that the robot teaching dialog is not available:
a
Click
Disconnect
in the RapidFire main software window.
b
Click
Stop
in the RapidFire Control Panel.
c
Close the RapidFire software.
d
Open the
RapidFire.cfg
file in Notepad and change
CONFIGURE_BENCHBOT=0;
.
4
Save and close the file and restart the RapidFire software.
Verifying the system operation
1
Set the temperature to 3 C and load the stacks with some plates of the type the customer is
using.
2
Verify the enclosure has cooled.
3
Run a batch file that uses the stacker and barcode reader.
4
Ensure the robot can access all stacks, place plates on the stage at the swap position and
barcode reader stage correctly.
Restore the system
1
Install the new rear panel that accommodates the TEC control.
2
Reinstall the side door (if removed) and the right-front door.