Instrument Installation Checklist
Revision:
A.00
, Issued:
January
17
, 20
21
Document part number:
D0007550
Page ____ of ____
© Agilent Technologies, Inc. 20
21
6
Route the pump outlet tube to a waste receptacle.
Installing the TEC controller
Unpack the controller and connect the following cables to the controller routed under the
deck:
Power (P/N 8121-2253)
Heater
TEC
Condensation pump (labeled Pump)
Temperature sensor
Serial (on the controller) to USB (on the computer) (P/N 8121-2109)
Connect the power cable to the upper power strip.
Turn on the TEC controller power switch located above the power cord connection.
Secure the TEC controller to the frame with M5 screws (P/N 0515-0387) and a T-25 driver.
Installing the front door to the enclosure
1
Pull out the drawer and partially install two lower M4 x 8-mm pan-head screws (P/N 0515-
2113, located in the door shipping box).
Figure 20 Partially installed attachment screws for the front door