Instrument Installation Checklist
Revision:
A.00
, Issued:
January
17
, 20
21
Document part number:
D0007550
Page ____ of ____
© Agilent Technologies, Inc. 20
21
Figure 7 Spacer edges aligned with the mounting plate
f
Lower the mounting plate onto the correction spacer by loosening the adjuster screw.
g
Install the mounting screw and remove the adjustment screw.
h
Remove the corresponding rear mounting screw and replace the O-ring and O-ring cup
with the second lean correction spacer.
Note: The weight of the drawer should be enough to lift up the mounting plate so you don't have
to use the adjuster to remove the rear O-ring and O-ring cup.
i
Replace the mounting screw.
j
Verify the change in tilt by checking the elevator to enclosure wall clearance.
Routing and connecting the enclosure cables
1
Feed the ribbon cables and temperature sensor cable through the cable port of the enclosure.
2
Using the M3 x 6-mm screw and T-10 driver, fasten the cable arm bracket to the elevator
mounting plate. Note that a single screw is used in the center hole of the bracket.
Figure 8 Alignment of the cable arm bracket to the center mounting hole