Instrument Installation Checklist
Revision:
A.00
, Issued:
January
17
, 20
21
Document part number:
D0007550
Page ____ of ____
© Agilent Technologies, Inc. 20
21
Setting the swap teachpoint
1
Set the swap teachpoint:
a
In the RapidFire software, click
System Tools > Sipper Configuration
.
b
Click
Home
to home the sipper, then click
Swap
to move the stage to the accessible
position for the plate handler.
c
Disable the robot motors and ensure the stage jig is correctly installed in the robot
grippers.
d
Move the robot so that the jig rests on the plate stage. Make sure it is flat and snug
against the side that holds the matrix bottles. You might have to move the x-axis of the
stage so that the jig is against the side that holds the matrix bottles. If you do move the
x-axis of the stage, click
Set swap X here
.
e
Close the doors and engage the motors.
f
Select
Swap
from the
Robot Teachpoint
list, and click
Teach at current Y,Z,Theta
.