Agilent B1500A User’s Guide, Edition 10
2- 33
Introduction
Measurement Units
WGFMU - Waveform Generator/Fast Measurement
Unit
This section describes typical specification of the waveform generator/fast
measurement unit (WGFMU) and the remote-sense and switch unit (RSU) for
Agilent B1500A. Maximum five modules can be installed in one mainframe.
•
Number of channels: 2 channels per module
•
Function:
Voltage output and current or voltage sampling measurement with minimum
sampling interval of 5 ns
•
Current measurement range: 1
A, 10
A, 100
A, 1 mA, or 10 mA
•
Operation mode: PG mode, Fast IV mode, or DC mode
•
PG mode:
ALWG (arbitrary linear waveform generator) voltage output and voltage
measurement (VFVM).
Output level:
-
5 V to 5 V for open load or
-
2.5 V to 2.5 V for 50
load
Minimum pulse width: 100 ns
•
Fast IV mode:
ALWG voltage output and current or voltage measurement (VFIM or VFVM).
Output level: 0 to
-
10 V, 0 to 10 V, or
-
5 V to 5 V
Minimum pulse width: 300 ns
•
DC mode:
DC voltage output and current or voltage measurement (VFIM or VFVM).
Output level: 0 to
-
10 V, 0 to 10 V, or
-
5 V to 5 V
Figure 2-8
WGFMU Simplified Circuit Diagram
50
7
V
PG mode
A
V
ALWG
Ch1
RSU output
Fast IV mode
50
7
V
PG mode
A
V
ALWG
Ch2
RSU output
Fast IV mode
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...