Agilent B1500A User’s Guide, Edition 10
2- 15
Introduction
Rear View
In the EasyEXPERT operation mode, the SPGU outputs can be specified by the
PG number. PG1 and PG2 are assigned to the Output 1 and Output 2 terminals
of the SPGU at the lowest position. And the following PG numbers are assigned
to the outputs of the upper SPGUs in sequence. Use the PG number label
furnished with the B1500A or equivalent to identify the SPGU channel number.
C. WGFMU module
The waveform generator/fast measurement unit (WGFMU) has the Ch 1 and Ch
2 terminals and the following terminals to synchronize the operation between
WGFMUs or with an external equipment.
Sync Out: Connect to the Sync In of the slave WGFMU.
Sync In: Connect to the Sync Out of the master WGFMU.
Trig Out: Connect to the trigger input terminal of the equipment synchronized
with the WGFMU output.
CAUTION
Connect the Sync Out/In, and Trig Out terminals to the specified terminal properly.
Connecting to the other terminal may result in damage to the WGFMU.
Connect the connection cable between the WGFMU and the RSU (remote-sense and
switch unit) before turning the B1500A on. Changing the cable connection in the
power on state may damage the WGFMU and RSU. See Agilent B1530A
User’s
Guide
for cable connection and usage.
D. SMU module
The source/monitor unit (SMU) has two triaxial connectors, force and sense, for
the Kelvin connections. The HRSMU (high resolution SMU) also provides a
D-sub connector (To ASU terminal) which is used to connect the dedicated ASU
(atto sense and switch unit). The HPSMU (high power SMU) occupies two slots.
In the EasyEXPERT operation mode, the SMUs can be specified by the SMU
number. SMU1 is assigned to the SMU at the lowest position. And the following
SMU numbers are assigned to the upper SMUs in sequence. Use the SMU
number label furnished with the B1500A or equivalent to identify the SMU
number.
NOTE
To use the ASU, connect the connection cables between the HRSMU and the
dedicated ASU before turning the B1500A on.
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...