Agilent B1500A User’s Guide, Edition 10
3- 41
Installation
Connecting Measurement Devices
Using Test Fixture
1. Press the
Stop
key to set the module output off.
2. Select a proper socket module for your DUT, then set the socket module on the
test fixture.
3. Mount your DUT on the socket module.
4. Connect between the socket module and the test fixture by using the proper test
leads.
5. Close the lid of the test fixture.
To force more than ± 42 V, close the lid of the test fixture. Otherwise, the
interlock function will stop Agilent B1500A output.
To connect between the socket module and the test fixture, you can use test leads
that have the following terminals:
•
Miniature banana-to-miniature banana
•
Miniature banana-to-pin plug
•
Miniature banana-to-miniature clip
WARNING
Make sure that the lid is closed properly before starting measurement. Do not
perform the measurement when a test lead is protruding from the test fixture
and the lid.
CAUTION
Do not connect or disconnect your DUT while Agilent B1500A is forcing voltage or
current. Otherwise, your DUT may be damaged.
CAUTION
Do not touch the terminals of the test leads. Oil, perspiration, and dirt prevent good
electrical contact, deteriorate insulation, and degrade measurement accuracy.
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...