Hardware
SMU Common Specifications
•
Output terminal/connection
Dual triaxial connector, Kelvin (remote sensing)
•
Voltage/current compliance (limiting)
The SMU can limit output voltage or current to prevent damaging the device
under test.
•
Voltage compliance:
0 V to ± 100 V (MPSMU, HRSMU)
0 V to ± 200 V (HPSMU)
•
Current compliance:
± 10 fA to ± 100 mA (HRSMU with ASU)
± 100 fA to ± 100 mA (HRSMU)
± 1 pA to ± 100 mA (MPSMU)
± 1 pA to ± 1 A (HPSMU)
•
Compliance accuracy:
Same as the current or voltage set accuracy.
•
Pulsed measurement
•
Width: 500
s to 2 s
•
Period: 5 ms to 5 s
Period
Width+2 ms (when Width
100 ms)
Period
Width+10 ms (when Width > 100 ms)
•
Resolution: 100
s
•
Pulse setting accuracy (fixed measurement range), supplemental information
•
Width: 0.5 % + 50
s
•
Period: 0.5 % + 100
s