![Agilent Technologies B1500A Скачать руководство пользователя страница 126](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697126.webp)
3- 52
Agilent B1500A User’s Guide, Edition 10
Installation
Capacitance Compensation When Using Switching Matrix
2. Measure the R, L, C values of the for C2H (DATA05), C2L (DATA06),
C3H (DATA07), and C3L (DATA08) by using the MFCMU. See “To obtain
compensation coefficients” on page 3-53.
After the measurements, calculate the per meter value of the R, L, C, and record
them into the following table.
3. Launch a text editor and open the compensation data file created at step 1.
Change the R, L, C values of the C2H (DATA05), C2L (DATA06),
C3H (DATA07), and C3L (DATA08) lines to the values recorded at step 2.
Do not change any other lines.
4. Save the file as your compensation data file (e.g. C:\temp\my_env_1.txt).
To perform the capacitance compensation, enter the file name of the
compensation data file to the User Compensation Data File field of the Extended
Configuration dialog box, and execute the measurement.
Before starting measurement, perform the CMU calibration. For details, see
Agilent EasyEXPERT
User’s Guide
.
Compensation
coefficients
Explanation
R (
)
L (H)
C (F)
C2H
(DATA05)
C2L
(DATA06)
C3H
(DATA07)
C3L
(DATA08)
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...