3- 26
Agilent B1500A User’s Guide, Edition 10
Installation
Installing Accessories
To Interconnect SPGUs
If multiple SPGUs have been installed, connect the SPGUs by using the following
parts. See Figure 3-7.
•
SPGU-SPGU connection cable16493Q, (number of SPGUs
-
1)
2 ea.
•
SMA open cap, 4 ea.
Figure 3-7
SPGU Module Connection Example, Four Modules
If multiple SPGUs have been installed, connect the SPGUs as shown in Figure 3-7.
This connection makes it possible to perform the synchronous output by the
multiple SPGU channels. In this figure, the master SPGU is the module which
outputs the reference signal and the slave SPGUs are the modules which follow to
the reference signal.
The Trig Out terminal must be connected to the trigger input terminal of the external
equipment to make the synchronized operation of it.
CAUTION
Connect the Ref Out/In, Sync Out/In, and Trig Out terminals to the specified
terminal properly. Connecting to the other terminal may result in damage to the
SPGU.
Master SPGU
Slave SPGU
Trig Out
Ref Out
Ref In
Sync Out
Sync In
Output 1
Output 2
16493Q SMA-SMA cable
SMA open cap
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...