![Agilent Technologies B1500A Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697059.webp)
Agilent B1500A User’s Guide, Edition 10
2- 19
Introduction
Measurement Units
HPSMU - High Power SMU
This section describes typical specification of the high power source/monitor unit
(HPSMU) for Agilent B1500A. Maximum four modules can be installed in one
mainframe.
•
Maximum voltage, current, output power: ± 200 V, ± 1 A, 20 W
•
Minimum range: 2 V, 1 nA
•
Output/measurement value and resolution: see Table 2-1 to Table 2-4.
Figure 2-3
HPSMU Output and Measurement Ranges
9ROWDJH9
&XUUHQWP$