![Agilent Technologies Agilent 7890 Series Скачать руководство пользователя страница 94](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996094.webp)
94
Maintaining Your GC
6
Maintaining the Purged Packed Inlet
To Install a Packed Glass Column
1
Gather the following:
•
9/16- inch wrench
•
Two 1/4- inch brass nuts (See
•
Two 1/4- inch Vespel/graphite ferrules
•
Lint- free gloves
2
Load the
and wait for the GC to
become ready.
3
Assemble a brass nut and Vespel/graphite ferrule on each
end of the column.
Glass columns must be simultaneously inserted into the
inlet and detector and installed parallel to the oven door.
When conditioning the column, do not attach the column
to the detector.
4
If conditioning the column, insert the column into the
purged packed inlet until it bottoms. Withdraw the
column 1 to 2 mm. Finger- tighten the inlet column nut.
(See
“To Condition a Packed Column”
WA R N I N G
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
C A U T I O N
Overtightening the column nut or forcing it to bottom in either the
inlet or detector may shatter the column.
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...