![Agilent Technologies Agilent 7890 Series Скачать руководство пользователя страница 346](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996346.webp)
346
Maintaining Your GC
19
Maintaining a Valve
To Install the Upper Valve Box
1
Gather the following:
•
T- 20 Torx screwdriver
•
3- mm hex key wrench
•
Flathead screwdriver
2
Verify that all valve rotors are in the full counterclockwise
position (valve
Off
).
3
For each actuator that mates with a newly installed valve:
a
Loosen the adjustment set screw.
b
Locate the rotor adjustment shaft on top of the
actuator. Use a screwdriver to rotate the valve rotor
counterclockwise until it stops.
4
Locate the two half- moon cutouts at the bottom back of
the upper valve box. Place the upper valve box on top of
the lower valve assembly, routing the heater/sensor wires
through the cutouts. Secure with two T- 20 mounting
screws.
5
Push each coupling/shaft assembly downward with a
flathead screwdriver until the slot on the coupling
engages the rotor index pin.
If the coupling and valve do not engage, check that both
are fully counterclockwise and try again. If necessary,
turn the shaft slightly to engage the coupling.
Rotor adjustment
slotted shaft
Adjustment set screw
Rotor adjustment shaft
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...