![Agilent Technologies Agilent 7890 Series Скачать руководство пользователя страница 229](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996229.webp)
Maintaining the FID
11
Maintaining Your GC
229
To Bakeout the FID
1
Bakeout the FID with the column installed or uninstalled.
If uninstalled, gather the following (see
•
Capillary adapter (adaptable FID only)
•
Column nut
•
No- hole ferrule
2
Load the
and wait for the GC to
become ready.
3
If the column is uninstalled, plug the detector connection
with the capillary adapter, column nut, and no- hole
ferrule.
Maintain inert carrier gas flow through the column, or
remove the column from the GC.
4
Set the detector temperature at 350 to 375 °C.
5
Set normal operating flows.
6
Light the FID flame.
7
Set the oven temperature to 250 °C or 25 °C above the
normal maximum operating temperature. Do not exceed
the column’s temperature limit.
WA R N I N G
If using hydrogen as a carrier gas, turn off the hydrogen supply
and cap the end of the column to prevent an oven explosion.
WA R N I N G
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...