![Agilent Technologies Agilent 7890 Series Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996090.webp)
90
Maintaining Your GC
6
Maintaining the Purged Packed Inlet
To Install a Packed Metal Column
1
Gather the following:
•
7/16- inch, 9/16- inch, and 1/2- inch wrenches
•
Lint- free gloves
2
Load the
and wait for the GC to
become ready.
3
Prepare the packed metal column. (See
Ferrules on a Packed Metal Column”
4
Install the 1/8- inch or 1/4- inch packed column inlet
adapter, if necessary. (See
5
Attach the column to the inlet adapter. Finger- tighten the
nut.
6
Tighten the nut an additional 1/4 turn with a wrench (for
a 1/8- inch column) or an additional 3/4 turn (for a
1/4- inch column).
Use two wrenches, one on the column nut and the other
on the adapter, to prevent the adapter from rotating.
7
Press
[Config]
,
[Column 1]
or
[Column 2]
,
enter
0.00
in either
Length
or
Diameter
, and identify the inlet and detector to
which the column is attached.
8
Condition the column, if necessary. (See
WA R N I N G
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
WA R N I N G
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...