50
Maintaining Your GC
5
Maintaining the Split/Splitless Inlet
9
Position the column so it extends 4 to 6 mm above the
end of the ferrule. Slide the septum up the column to
hold the column nut at this position.
10
Thread the column nut into the inlet but do not tighten.
11
Adjust the column position so that the septum contacts
the bottom of the column nut. Finger- tighten the column
nut until it begins to grip the column.
12
Tighten the column nut an additional 1/4 to 1/2 turn
with a wrench so that the column cannot be pulled from
the fitting with gentle pressure.
13
Configure the new column. If using the optional barcode
scanner, scan the column directly into your data system
software configuration, then download the changes to the
GC. `
14
Condition the column per the manufacturer’s
recommendation. (See
To Condition a Capillary Column
.)
15
Install the column into the detector. See:
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the uECD
To install a Capillary Column in the FPD Plus
•
To Install a Capillary Column Adapter in the FPD
4
%
6 mm
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...