![Agilent Technologies Agilent 7890 Series Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996014.webp)
14
Maintaining Your GC
1
About Maintaining the GC
•
The filament in the TCD will be damaged if exposed to
air while hot. To protect the filament, turn it
Off
.
•
Set all detector flows to
Off
.
Once the zones reach < 70 °C, you can perform general GC
maintenance.
Inlet Maintenance Method
This method prepares the inlet for maintenance while
leaving the detector at operating temperature.
•
Set the oven temperature to
35 °C
. This allows the oven
fan to assist cooling.
•
Set all inlet temperatures to
Off
and set inlet gas
pressures to
0.0
.
•
If performing column maintenance, remember to wait
for the oven and column to cool down before turning
off column carrier gas flow at the source. Also
remember to cap both ends of the column to keep air
out once it is removed.
•
If you are not performing column maintenance, keep
inert carrier gas (helium or nitrogen) flowing to protect
the column.
•
Maintain all temperature setpoints for installed detectors,
if desired.
•
The filament in the TCD will be damaged if exposed to
air while hot. To protect the filament, turn it
Off
.
Once the zones reach < 70 °C, you can perform general GC
maintenance.
WA R N I N G
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
WA R N I N G
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...