709
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_fer.fm
Frame Error Rate Measurement Troubleshooting
Frame Error Rate Measurement Troubleshooting
Possible Setup Issues
IMPORTANT
If the Access Probe Power measurement is currently selected and the Trigger Arm setting is
Continuous, all other measurements are disabled. Use the
command to set the trigger arm state to single and retry this measurement.
During remote operation of this measurement, you should configure the Trigger Arm to Single. See
“SETup:CFERror:CONTinuous” on page 526
. If single triggering is used during manual operation, press the
START SINGLE key to trigger each measurement.
The test set receiver’s expected signal amplitude and channel number (or frequency) must be set up correctly
for accurate measurements. This is normally done automatically (see
) but can
also be done manually (see
“Receiver Control” on page 194
) .
A call must be connected with a loopback service option selected (see
) to perform
frame error rate measurements.
Frame error rate measurements require an accurate Amplitude Offset value because of the low levels
transmitted during sensitivity measurements. Use the
command to set offset
values.
Interpreting Integrity Indicator values
See
“Integrity Indicator” on page 222
.
Related Topics
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...