737
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_err_msg_pos100.fm
+100 t199 Error Message Descriptions
+113 Internal error; <VI NAME> forced inactive
Indicates that a VI is inactivated when not executed.
<VI NAME> includes : “IntVmVI”,”GprsBsEmulVI”,
“GsmComBsEmulVI”,“GSMFixedVI”, “MiscVI”,
“GSMSacchMriVI”, “IQSelfCalVI1”, “IQSelfCalVI2”,
“TA RevisionVI”, “TdmaTaBsEmulVI”.
+114 Internal error; <VI NAME> not responding
Indicates that a VI has not been instantiated or the
state is not available.
<VI NAME> includes : “IntVmVI”,”GprsBsEmulVI”,
“GsmComBsEmulVI”, “GSMFixedVI”, “MiscVI”,
“GSMSacchMriVI”, “IQSelfCalVI1”, “IQSelfCalVI2”,
“TA RevisionVI”, “TdmaTaBsEmulVI” ,
“RfHwAccessVI”.
+115 Internal error; <VI NAME> inactive
Indicates that a VI has been made inactive.
<VI NAME> includes : “IntVmVI”, “GSMFixedVI”,
“MiscVI”, “GSMSacchMriVI”, “IQSelfCalVI1”,
“IQSelfCalVI2”, “TA RevisionVI”.
+120 Warning; Receiver over range due to requested
settings
Indicates the total received power (a combination of
received power and amplitude offset) is above the
range of the test set.
+121 Warning; Receiver under range due to requested
settings
Indicates the total received power (a combination of
received power and amplitude offset) is below the
range of the test set.
+122 Warning; Reference out of lock
Indicates the test set’s internal reference is out of lock.
+123 Warning; Duplicate RF IN/OUT Amplitude Offset
Frequency entry. First frequency entry in RF IN/OUT
Amplitude Offset table will be used.
Indicates that an amplitude offset value for that
frequency has already been entered. The test set will
use the amplitude offset value entered first.
+124 Warning; Source over range due to requested
settings
The setting selected for total transmit power (a
combination of cell power, awgn power, amplitude
offset) is above the test set’s range.
+125 Warning; Source under range due to requested
settings
The setting selected for total transmit power (a
combination of cell power, awgn power, amplitude
offset) is below the test set’s range.
+130 Configuration error; Unable to switch to indicated
Test Application
The test application selected is not available for this
test set.
+131 Configuration error; Current Test Application is
not licensed (License = N). Select another Test
Application
The test application selected is not licensed. Select
another revision or test application that is licensed
(License = L).
Error Message
Description
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...