Programming a SINAD Measurement
Programming a SINAD Measurement
This section provides an example of how to make SINAD measurement via GPIB. It is patterned after
TIA/EIA IS-137, Analog RF Sensitivity (section 2.3.1.1).
SINAD and Distortion measurements are affected by the same STATe and INITiate commands, so both
measurements are enabled and triggered at the same time. However, measurement results are queried
separately. See
“Audio Analyzer Measurement Description” on page 38
.
The following procedure assumes that the mobile’s audio output is connected to the AUDIO IN connectors and
the mobile is on an analog voice channel (AVC) that is being modulated with a 1004 Hz tone at
±
8 kHz peak
deviation. It is also assumed that the Cell Power from the test set is set to a minimum usable level (typically
about -116 dBm).
1. Configure audio analyzer measurement parameters using the SETup subsystem.
2. Start the audio analyzer measurement using the INITiate subsystem.
3. Use the FETCh? subsystem to obtain audio analyzer measurement results.
Programming Example
10 OUTPUT 714;”SETup:AFANalyzer:CONTinuous OFF” !Set the audio analyzer
20 !measurements to single trigger mode.
30 OUTPUT 714;”SETup:AFANalyzer:PEAK:VOLTage 3V” !Set the Expected Amplitude level for
40 !audio input voltage in Vpeak.
50 OUTPUT 714;”SETup:AFANalyzer:SDIStortion:STATe ON” !Turn on the SINAD and
60 !distortion measurements.
70 OUTPUT 714;”SETup:AFANalyzer:SDIStortion:FREQuency 1004 HZ” !Specify the audio
80 !frequency to use for the measurements.
90 OUTPUT 714;”SETup:AFANalyzer:FILTer CMESsage” !Select the c-message filter.
100 OUTPUT 714;”SETup:AFANalyzer:DEMPhasis:STATe OFF” !Turn off de-emphasis.
110 OUTPUT 714;”SETup:AFANalyzer:EXPandor:STATe OFF” !Turn off the expandor.
120 OUTPUT 714;”SETup:AFANalyzer:TIMeout 3S” !Set a timeout value of 3 seconds
130 !in case the measurement cannot be made.
140 OUTPUT 714;”INITiate:AFANalyzer” !Start the AF Analyzer measurement.
150 OUTPUT 714;”FETCh:AFANalyzer:INTegrity?” !Query the integrity indicator to
160 !verify that a reliable measurement was made.
170 ENTER 714;Integrity !Enter the returned value into a variable for comparison
180 !with possible integrity indicator values (not shown here).
190 IF Integrity=0 THEN !Only fetch measurement result if integrity indicator is 0.
200 OUTPUT 714;”FETCh:AFANalyzer:SINAD?” !Fetch the SINAD result.
210 ENTER 714;Sinad !Enter the returned value into a variable.
220 END IF
230 END
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...