224
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_prog_gen_integrity.fm
Integrity Indicator
17
Can not Correlate: Indicates that the test sets internally generated reference signal does not
correlate with the received signal.
Some conditions that could cause this integrity indicator result include the following:
• an input signal that is corrupted
• the input signal is extremely distorted
• the input signal is off by more than 10 ms.
• the frequency of the input signal deviates more than allowed
• something is wrong with the long-code mask coming from the mobile ID
18
Frequency Out Of Range: Indicates that a Channel Power Measurement was attempted at a
frequency for which there is no calibration data. The test set display will indicate 4 dashes.
19
Uncalibrated Due To Temperature: Indicates that the current temperature of the test set is
different than the calibration data temperature by more than ±10 degrees C, when attempting a
Channel Power Measurement.
20
Potential Receiver Saturation: Indicates that an input signal immediately prior to the measured
input was high enough to potentially saturate the receiver hardware.
21
Parameter Error: Indicates that a measurement setup parameter has been set in a way that gives
invalid measurement results.
22
Unsupported Configuration: Indicates that some parameter, other than a measurement setup
parameter has been set so that it causes an invalid measurement result. Typically this would be a
base station emulator parameter.
23
Call Processing Operation Failed: Indicates that a call processing operation, (base station
emulator) needed in the course of making the measurement could not complete.
Table 13.
Integrity
Indicator
Number
Integrity Indicator Message
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...