713
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_gpower.fm
Gated Power Measurement Troubleshooting
Gated Power Measurement Troubleshooting
Possible Setup Issues
IMPORTANT
If the Access Probe Power measurement is currently selected and the Trigger Arm setting is
Continuous, all other measurements are disabled. Use the
command to set the trigger arm state to single and retry this measurement.
During remote operation of this measurement, you should configure the Trigger Arm to Single. See
. If single triggering is used during manual operation, press the START
SINGLE key to trigger each measurement.
The test set receiver’s expected signal amplitude and channel number (or frequency) must be set up correctly
for accurate measurements. This is normally done automatically (see
) but can
also be done manually (see
“Receiver Control” on page 194
) .
The following settings are required for the gated power measurement:
• Operating Mode must be Active Cell
• System Type must be IS-95 or IS-2000 with Radio Config set to (Fwd1, Rvs1) or (Fwd2, Rvs2)
• FCH Service Option Setup must be set to SO2 or SO9 (loopback service option)
• Traffic Data Rate must be set to Eighth.
The input signal must conform to
.
Related Topics
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...