526
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_cfer.fm
SETup:CFERror
SETup:CFERror:CONFidence:REQuirement[:RATio]
SETup:CFERror:CONTinuous
SETup:CFERror:COUNt[:MAXimum]
Function
Sets/queries the FER (frame error rate) measurement’s FER requirement. This value specifies
the percentage of frame errors used in determining confidence limit pass/fail results.
Setting
Range: 0.10 to 5.0
Query
Range: 0.10 to 5.0
*RST Setting
1.00 %
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE:REQUIREMENT 2 PCT”
!sets the confidence limit requirement to 2.00 percent.
Function
Sets/queries the trigger arm state for the FER (frame error rate) measurement. The trigger arm
state determines whether the waveform quality measurement will be made once then stop
(0|OFF (single)), or automatically rearm upon completion of one measurement and repeat the
process (1|ON (continuous)).
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONTINUOUS OFF
”
!Sets the FER measurement to single trigger mode.
Function
The setting form of this command sets the maximum frame count parameter for frame error rate
measurements.
If the confidence level parameter is off, the FER (frame error rate) measurement will continue
until the maximum frame count is reached. If the confidence level parameter is on and the FER
measurement reaches max frames, the FER measurement will return the string value “Max
Frames” (see
The query form returns the maximum frame count parameter.
Setting
Range: 25 to 10,000,000
Query
Range: 25 to 10,000,000
*RST Setting
10,000
Programming Example
OUTPUT 714;"SETUP:CFERror:COUNT 10000”
!sets the maximum frames value to 10,000.
Содержание 8960
Страница 14: ...14 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide cdma2000_titlepage fm ...
Страница 33: ...33 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_measurements fm Measurements Measurements ...
Страница 105: ...105 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_general_programming fm Programming Programming ...
Страница 247: ...247 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_hpib_command_syntax fm GPIB Commands ...
Страница 654: ...654 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_hpib_common fm IEEE 488 2 Common Commands ...
Страница 655: ...655 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_manual_operation fm Manual Operation ...
Страница 701: ...701 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_error_messages fm Troubleshooting ...
Страница 759: ...759 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters ch_configure fm Configuration Configuration ...
Страница 780: ...780 S Hp8960 E1962B CDMA 2000 Pi Release Reference Guide Chapters cdma2000_conf_name fm Test Application Name ...