3-21
Agilent 54621A/21D Performance Test Record
Agilent 54621A/54621D
Serial No. ______________________________________
Test by _____________________________
Test Interval ____________________________________
Work Order No. ______________________
Recommended Next Testing ________________________
Temperature ____________
Threshold
Specification
Limits
Ch D7-D0
Ch D15-D8
Accuracy Test
5 V - 250 mV
4.750 V
________
________
(100 mV + 3% of
threshold setting)
5 V + 250 mV
5.250 V
________
________
-5 V - 250 mV
-5.250 V
________
________
-5 V + 250 mV
-4.750 V
________
________
0 V - 100 mV
-100 mV
________
________
0 V + 100 mV
100 mV
________
________
Voltage Measurement Accuracy
Range
Power Supply Setting
Test Limits
Channel 1 Channel 2 Channel 3 Channel 4
5 V/Div
35 V
34.04 V to 35.96 V
________
________ ________ ________
2 V/Div
14 V
13.616 V to 14.384 V
________
________ ________ ________
1 V/Div
7 V
6.808 V to 7.192 V
________
________ ________ ________
500 mV/Div
3.5 V
3.404 V to 3.596 V
________
________ ________ ________
200 mV/Div
1.4 V
1.3616 V to 1.4384 V
________
________ ________ ________
100 mV/Div
700 mV
680.8 mV to 719.2 mV
________
________ ________ ________
50 mV/Div
350 mV
340.4 mV to 359.6 mV
________
________ ________ ________
20 mV/Div
140 mV
136.16 mV to 143.84 mV
________
________ ________ ________
10 mV/Div
70 mV
68.08 mV to 71.92 mV
________
________ ________ ________
5 mV/Div
35 mV
34.04 mV to 35.96 mV
________
________ ________ ________
2 mV/Div
14 mV
13.616 mV to 14.384 mV
________
________ ________ ________
1 mV/Div
7 mV
6.616 mV to 7.384 mV
________
________ ________ ________
Bandwidth
Test Limits
Channel 1 Channel 2
3 dB at 60 MHz
________
________
Horizontal
∆
t and 1/
∆
t Accuracy
Generator Setting
Test Limits
Results
Frequency
10 kHz
9.98 kHz to 10.02 kHz
________
Period
100
µ
s
99.79
µ
s to 100.2
µ
s
________
Frequency
1 MHz
997.9 kHz to 1.002 MHz
________
Period
1
µ
s
997.9 ns to 1.002
µ
s
________
Frequency
50 MHz
49.77 MHz to 50.23 MHz ________
Period
20 ns
19.91 ns to 20.09 ns
________
Trigger Sensitivity
Test Limits
Channel 1 Channel 2
Internal trigger
25 MHz at 0.35 division
________
________
60 MHz at 0.35 division
________
________
Ext
External trigger
25 MHz at <75 mV
p-p
________
60 MHZ at <150 mV
p-p
________
service.book Page 21 Wednesday, December 18, 2002 8:35 AM
Содержание 54621A
Страница 4: ...iv service book Page iv Wednesday December 18 2002 8 35 AM ...
Страница 7: ...1 General Information service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 24: ...1 18 service book Page 18 Wednesday December 18 2002 8 35 AM ...
Страница 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 44: ...2 20 service book Page 20 Wednesday December 18 2002 8 35 AM ...
Страница 45: ...3 Testing Performance service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 66: ...3 22 service book Page 22 Wednesday December 18 2002 8 35 AM ...
Страница 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 76: ...4 10 service book Page 10 Wednesday December 18 2002 8 35 AM ...
Страница 77: ...5 Troubleshooting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 90: ...5 14 service book Page 14 Wednesday December 18 2002 8 35 AM ...
Страница 91: ...6 Replacing Assemblies service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 107: ...7 Replaceable Parts service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 114: ...7 8 service book Page 8 Wednesday December 18 2002 8 35 AM ...