1-11
General Information
Vertical System: Analog Channels (continued)
Vertical System: Analog Channels (continued)
ESD Tolerance
±2 kV
Noise Peak-to-Peak
2% full scale or 1 mV, whichever is greater
Common Mode Rejection Ratio
20 dB @ 50 MHz
DC Vertical Gain Accuracy*
1
±2.0% full scale
DC Vertical Offset Accuracy
< 200 mV/div: ±0.1 div ±1.0 mV ±0.5% offset
≥
200 mV/div: ±0.1 div ±1.0 mV ±1.5% offset value
Single Cursor Accuracy
1
±{DC Vertical Gain Ac DC Vertical Offset
Ac 0.2% full scale (~1/2 LSB) }
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0%(80mV) + 0.1 (10 mV) + 1.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 3.78 mV
Dual Cursor Accuracy*
1
±{DC Vertical Gain Ac 0.4% full scale (~1 LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0%(80 mV) + 0.4%(80 mV)} = ±1.92 mV
1
1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.
Vertical System: Digital Channels (54621D and 54622D only)
Number of Channels
16 Digital – labeled D15 – D0
Threshold Groupings
Pod 1: D7 - D0
Pod 2: D15 - D8
Threshold Selections
TTL, CMOS, ECL, user-definable (selectable by pod)
User-Defined Threshold Range
±8.0 V in 10 mV increments
Maximum Input
Voltage
±40 V peak CAT I
Threshold Accuracy*
±(100 mV + 3% of threshold setting)
Input Dynamic Range
±10 V about threshold
Minimum Input Voltage Swing
500 mV peak-to-peak
Input Capacitance
~ 8 pF
Input Resistance
100 k
Ω
±2% at probe tip
Channel-to-Channel Skew
2 ns typical, 3 ns maximum
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 °C from firmware calibration temperature.
service.book Page 11 Wednesday, December 18, 2002 8:35 AM
Содержание 54621A
Страница 4: ...iv service book Page iv Wednesday December 18 2002 8 35 AM ...
Страница 7: ...1 General Information service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 24: ...1 18 service book Page 18 Wednesday December 18 2002 8 35 AM ...
Страница 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 44: ...2 20 service book Page 20 Wednesday December 18 2002 8 35 AM ...
Страница 45: ...3 Testing Performance service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 66: ...3 22 service book Page 22 Wednesday December 18 2002 8 35 AM ...
Страница 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 76: ...4 10 service book Page 10 Wednesday December 18 2002 8 35 AM ...
Страница 77: ...5 Troubleshooting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 90: ...5 14 service book Page 14 Wednesday December 18 2002 8 35 AM ...
Страница 91: ...6 Replacing Assemblies service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 107: ...7 Replaceable Parts service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 114: ...7 8 service book Page 8 Wednesday December 18 2002 8 35 AM ...