3- 12
Testing Performance
To verify voltage measurement accuracy
3
Use the BNC tee and cables to connect the oscilloscope calibrator
/power supply to both the oscilloscope and the multimeter.
4
Adjust the output so that the multimeter reading displays the first
Volts/div supply setting value in Table 3-5.
Wait a few seconds for the measurement to settle.
5
Press the
Y2
softkey, then position the Y2 cursor to the center of the
voltage trace using the Entry knob.
The
∆
Y value on the lower line of the display should be within the test limits of
Table 3-5. If a result is not within the test limits, see the “Troubleshooting”
chapter. Then return here.
6
Continue to check the voltage measurement accuracy with the
remaining Volts/div setting values in Table 3-5.
7
When you are finished checking all of the power supply setting values,
disconnect the power supply from the oscilloscope.
8
Repeat this procedure for Channels 2, 3, and 4, if applicable on your
oscilloscope model.
Figure 3-4
Using a Blocking Capacitor to Reduce Noise
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise may
be a factor. To eliminate the noise, use a BNC Tee, blocking capacitor, and BNC
shorting cap to shunt the noise to ground. See figure 3-4.
To oscilloscope input
BNC shorting
cap
To Power Supply or
Calibrator
Blocking
Capacitor
service.book Page 12 Wednesday, December 18, 2002 8:35 AM
Содержание 54621A
Страница 4: ...iv service book Page iv Wednesday December 18 2002 8 35 AM ...
Страница 7: ...1 General Information service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 24: ...1 18 service book Page 18 Wednesday December 18 2002 8 35 AM ...
Страница 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 44: ...2 20 service book Page 20 Wednesday December 18 2002 8 35 AM ...
Страница 45: ...3 Testing Performance service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 66: ...3 22 service book Page 22 Wednesday December 18 2002 8 35 AM ...
Страница 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 76: ...4 10 service book Page 10 Wednesday December 18 2002 8 35 AM ...
Страница 77: ...5 Troubleshooting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 90: ...5 14 service book Page 14 Wednesday December 18 2002 8 35 AM ...
Страница 91: ...6 Replacing Assemblies service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 107: ...7 Replaceable Parts service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 114: ...7 8 service book Page 8 Wednesday December 18 2002 8 35 AM ...