6- 8
Replacing Assemblies
To remove the display
To remove the display
1
Remove the front panel from the oscilloscope.
2
Disconnect the ribbon cable from the display board.
3
Using the T20 TORX driver, remove the four screws that hold the display
to the deck. Set these screws aside.
Figure 6-5
Removing the display
C A U T I O N
USE CORRECT SCREWS TO AVOID SHORTING SYSTEM BOARD !
When you re-install the display, you must re-insert the screws you removed in
this step. If you re-insert screws that are longer, they can short the system
board to ground.
service.book Page 8 Wednesday, December 18, 2002 8:35 AM
Содержание 54621A
Страница 4: ...iv service book Page iv Wednesday December 18 2002 8 35 AM ...
Страница 7: ...1 General Information service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 24: ...1 18 service book Page 18 Wednesday December 18 2002 8 35 AM ...
Страница 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 44: ...2 20 service book Page 20 Wednesday December 18 2002 8 35 AM ...
Страница 45: ...3 Testing Performance service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 66: ...3 22 service book Page 22 Wednesday December 18 2002 8 35 AM ...
Страница 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 76: ...4 10 service book Page 10 Wednesday December 18 2002 8 35 AM ...
Страница 77: ...5 Troubleshooting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 90: ...5 14 service book Page 14 Wednesday December 18 2002 8 35 AM ...
Страница 91: ...6 Replacing Assemblies service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 107: ...7 Replaceable Parts service book Page 1 Wednesday December 18 2002 8 35 AM ...
Страница 114: ...7 8 service book Page 8 Wednesday December 18 2002 8 35 AM ...