Computer Connection
3
User Manual
23
Once you have removed the top cover, set switch position 1 ON to enable
communication mode. In this mode, the ON switch positions for switches 2-8
are defined in
An example jumper setting for this switch is shown in
When switch 1 is set to the ON position, RS-232 port is activated. Switch 2 set
to ON indicates that handshaking is enabled. With handshake enabled, switch
3 set to OFF indicates software handshaking. Switch 4 set to ON enables parity
checking. Switch 5 set to OFF indicates odd parity.
Switches 6 through 8 combine to indicate baud rate. See
for switch
settings and their corresponding baud rates. Switches 6 and 8 set to ON and 7
set to OFF combine to form a RS-232 baud rate of 9600 as shown in
Table 3
RS-232 communication mode
Switch
ON Action
2
Handshake enabled
3
Hardware handshake (otherwise software handshake)
4
Parity enable
5
Even parity (otherwise odd parity)
6, 7, 8
Baud rate (see
)
Figure 5
Example jumper settings for RS-232
1 2 3 4 5
7
6
8
ON
OFF