5
Troubleshooting and Maintenance
280-DS Operator’s Manual
99
Cannot perform functions
within the software (e.g.,
create system/method, run
method, etc.).
User does not have proper
rights/access to desired
function.
•
Administrator must add desired
user to individual Vk groups to
grant access/privileges to specific
software functions via User
Administration (Tools > Options).
See “Adding Users to the
Application” on page 25.
User cannot log on to
software.
Incorrect password or
domain.
•
Verify the correct password has
been entered and the proper
domain is selected from the
software login screen.
Data not updating in
real-time or SIMULATED is
displayed as the serial
number of dissolution
apparatus.
Use Simulated Instruments
option enabled.
•
Disable the
Use Simulated
Instruments
option under Tools >
Simulated Instruments” on
page 92..
Symptom
Probable Cause
Suggested Solution
Содержание 280-DS
Страница 1: ...280 DS G7980AA Mechanical Qualification System Operator s Manual ...
Страница 5: ...280 DS Operator s Manual 5 1 Safety Notes Cautions and Warnings 6 ...
Страница 8: ...1 Safety 8 280 DS Operator s Manual This page was intentionally left blank except for this message ...
Страница 9: ...280 DS Operator s Manual 9 2 Introduction ...
Страница 58: ...4 Operating the 280 DS 58 280 DS Operator s Manual Figure 36 Paddle Basket Height ...
Страница 76: ...4 Operating the 280 DS 76 280 DS Operator s Manual Figure 54 Test Completion Screen Unsuccessful ...
Страница 95: ...280 DS Operator s Manual 95 5 Troubleshooting and Maintenance Preventive Maintenance 96 Troubleshooting 97 ...
Страница 101: ...280 DS Operator s Manual 101 6 Service and Warranty Service and Warranty Information 102 ...