4
Operating the 280-DS
280-DS Operator’s Manual
45
Selecting the Stirring Element
1
Select the applicable stirring element for the test and check the boxes next to
the positions you want to test.
2
Click
Next
to continue.
Figure 23.
Stirring Element and Vessel Positions
NOTE
If any of the Active Positions are unchecked, the software will
notify you that the dissolution apparatus qualification record will
not be updated since no positions are being tested.
For any test procedure executed with less than the complete set of
active vessels, FOR REFERENCE ONLY will appear on the final test