4
Operating the 280-DS
70
280-DS Operator’s Manual
Bath Temperature Measurement
The Bath Temperature Measurement page displays the reading of the 280-DS
temperature probe and the bath temperature reading of the dissolution
apparatus, if it is online.
1
If the dissolution apparatus is offline, then you must enter the bath
temperature as reported on its display. Although a successful test requires
reaching the target temperature, you can accept the result at anytime.
Figure 48.
Bath Temperature Measurement
2
After 30 seconds, the reading stabilizes. Click
Accept
to continue to the next
screen.
NOTE
If applicable, do not remove the bath temperature probe of the
dissolution apparatus from the water bath for this measurement.
Instead, remove one of the rear dissolution vessels for access to
the water bath.