4
Operating the 280-DS
280-DS Operator’s Manual
75
Test Completion
1
Upon completion of the test sequence:
•
If the method has the Update Qualification Record option selected
(
on page 38) and all of the parameters have
passed the specification, you can specify the next qualification due date.
•
If any of the tests failed, then the dissolution apparatus is marked Out of
Calibration and the old calibration due date cannot be updated.
•
If Update Qualification Record is not selected then the old qualification due
date remains and will be displayed for reference only.
Figure 53.
Test Completion Screen - Successful
2
Click
Complete
to update the physical parameters and display the Method
Completion screen. When the test is complete, the report is compiled and
displayed on screen. See “Test Reports” on page 82..