16900A, 16902A, and 16903A Service Guide
6
Replacing Assemblies
To remove and replace optional modules or filler panels
The following procedure guides you through installing,
removing, or replacing measurement modules in the logic
analysis system frame. Information on connecting specific
multi- card measurement modules can be found in the
Installation Guide that came with your frame. Measurement
modules with different model numbers may not be connected
together in multi- card (Master/Expander) modules unless
stated otherwise. See also, the service guide that came with
your measurement module.
1
Perform previous procedures:
•
“To prepare the instrument for disassembly"
2
Using a 1/4- inch hex socket, loosen the thumb screws on
the modules or filler panels.
3
Rotate the ejector levers from the locked position to the
open position.
4
Ensure all ejectors are fully open and out of the way.
5
Remove filler panels or slide the modules out of the logic
analysis system frame.
The following figure shows a 6- card frame. The
insertion/removal order would be the same for a 3- card
frame. A single- module configuration can be installed in
any available slot.
Locked position
Open position
Thumb screw
Содержание 16900A
Страница 1: ...Agilent Technologies Agilent 16900A 16902A and 16903A Logic Analysis System Service Guide...
Страница 6: ...6 16900A 16902A and 16903A Service Guide...
Страница 14: ...14 16900A 16902A and 16903A Service Guide 1 General Information...
Страница 18: ...18 16900A 16902A and 16903A Service Guide 2 Preparing for Use...
Страница 22: ...22 16900A 16902A and 16903A Service Guide 3 Testing Performance...
Страница 86: ...86 16900A 16902A and 16903A Service Guide 6 Replacing Assemblies...
Страница 102: ...102 16900A 16902A and 16903A Service Guide 7 Replaceable Parts...