![Agilent Technologies 16900A Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/agilent-technologies/16900a/16900a_service-manual_2867161012.webp)
12
16900A, 16902A, and 16903A Service Guide
1
General Information
Electrical and Operating Environment Characteristics
The following operating characteristics are not specifications,
but are typical operating characteristics for the Agilent
16900- series logic analyzers.
Electrical Characteristics
Power Requirements
Frame(16900/902A): 115/230 Vac +/- 20%, 48-66Hz, 1300W Max
Frame(16903A): 115/230 Vac +/- 20%, 48-66Hz, 900W Max
The line voltage is autodetected by the instrument.
CAT II (Line voltage in appliance and to wall outlet)
Pollution degree 2
Environmental Characteristics
For indoor use only.
Temperature
Instrument (16900A & 16902A): 0° to + 40° C (+32° to +104° F)
Instrument (16903A): 0° to + 50° C (+32° to +122° F)
Disk Media: 10° to + 40° C (+50° to +104° F)
Probes/cables: 0° to + 65° C (+32° to +149° F)
Reliability is enhanced when operating the instrument within a temperature of +20°C to +35°C
(+68°F to +95°F).
Altitude
3,000 m (10,000 ft)
Humidity
Relative humidity 8% to 80% at 40° C (104° F). Avoid sudden, extreme temperature changes which
could cause condensation on the circuit board.
Reliability is enhanced when operating the instrument within a noncondensing humidity range of
20% to 80%. Condensation within the instrument can cause poor operation or malfunction. Provide
protection against internal condensation.
Vibration
Operating: random vibration 0-500 Hz, 10 minutes per axis, 0.3 g (rms).
Non-operating: random vibration 0-500 Hz, 10 minutes per axis, 2.41g (rms); and swept sine resonant
search, 0-500 Hz, 0.75g, 5-minute resonant dwell at 4 resonances per axis.
Weight
Model Max
Net Max
Shipping
16900A 16 kg (35.2 lbs) 24.6 kg (54.2 lbs)
16902A 17.2 kg (37.8 lbs) 25.8 kg (56.8 lbs)
16903A 14.5 kg (32.0 lbs) 23.2 kg (23.2 lbs)
The weight of modules ordered with mainframes will add 0.9 kg (2.0 lb) per module
Storage
Store or ship the instrument in environments within the following limits:
Temperature -40°C to +75°C (-40°F to 167°F)
Humidity up to 90% at 65° C (149°F)
Altitude up to 3,000 meters (10,000 feet)
Protect the system from temperature extremes which cause condensation on the instrument.
Содержание 16900A
Страница 1: ...Agilent Technologies Agilent 16900A 16902A and 16903A Logic Analysis System Service Guide...
Страница 6: ...6 16900A 16902A and 16903A Service Guide...
Страница 14: ...14 16900A 16902A and 16903A Service Guide 1 General Information...
Страница 18: ...18 16900A 16902A and 16903A Service Guide 2 Preparing for Use...
Страница 22: ...22 16900A 16902A and 16903A Service Guide 3 Testing Performance...
Страница 86: ...86 16900A 16902A and 16903A Service Guide 6 Replacing Assemblies...
Страница 102: ...102 16900A 16902A and 16903A Service Guide 7 Replaceable Parts...