50
1141A and 1142A User’s Guide
3
Adjusting the Probe
Probe Adjustment
Figure 27
Signal to +Input
26
Adjust R14 (
LF Gain
) and C4 (
+ LF BANDWIDTH
) for the flattest pulse
top (see
on page 50). Again, adjust slowly and press
CLEAR DISPLAY
frequently to restart averaging.
Figure 28
R14 and C4 Adjustment
27
Carefully connect the probe to the test board in position shown in
the figure below (signal to both inputs).
Signal to Both Inputs