1141A and 1142A User’s Guide
3
Adjusting the Probe
Probe Adjustment
Figure 30
R13 Adjustment
NOTE
The minimum allowable amplitude is 212 mV
p-p
. Typical values will be between 275
and 325 mV
p-p
. the probe needs repair if the minimum cannot be reached.
Probe Reassembly
35
Disconnect the probe power cable at the rear panel of the 1142A
36
Remove the probe PC assembly from the test board. Be sure the
probe input connectors remain attached to the probe.
37
Remove the grounding screw and ground block from the PC
assembly.
38
Assemble the PC assembly in the top cover.
NOTE
The side of the PC assembly with the large hybrid is exposed when the assembly is
in the top cover. The figure below shows how the top cover, PC board, and ground
block fit together.