48
1141A and 1142A User’s Guide
3
Adjusting the Probe
Probe Adjustment
16
Carefully connect the input of the probe to the test board in the
position shown in the
Figure 24
Signal to +Input
17
Center adjustment R11,
HF CMRR
, as shown in
Figure 25
R11, HF CMRR Adjustment
18
Adjust R9,
HF GAIN
so the V
p- p
(1) measurement is the same as in
step 4, within ±0.5%. Make the adjustment slowly so the
oscilloscope display has time to react to signal averaging. Press
CLEAR DISPLAY
occasionally to restart averaging, which gives a
quicker indication of changes.
19
Carefully connect the probe to the test board in the position shown
in