Service
4
Troubleshooting
1141A and 1142A User’s Guide
65
Troubleshooting
Use the following paragraphs to assist in troubleshooting problems
with the 1141A/1142A Differential Probe.
Probe Troubleshooting
To troubleshoot the probe:
1
Apply a known signal to the input of the probe.
2
Check for an identical output at the output coax to the cable. This
connection is the one soldered to the PC board. If the probe output
cable is not terminated, or the coax is open, the output signal will
be about twice the amplitude of the input signal.
3
If the signal is incorrect, check the power supply voltages from the
1142A Probe Control and Power Module. Use the cable diagram on
the next page.
4
Troubleshoot the cable with an ohmmeter. Use the cable diagram
on the next page.
Probe Control and Power
Module Troubleshooting
The circuitry consists of simple power supplies, operational
amplifiers, and TTL. Use conventional troubleshooting techniques. A
complete parts list, component locater, and schematics are provided
later in this chapter.