74
4454C EN 2013-05-14
Specifications
CR 35-X D
IGITIZER
Warming-up time
• Cold start
fully operational after max. 30 min.
• Warm start
fully operational after self-test
if not switched off for more than 3 min., after 30
min. operation
Physical emissions
Noise emission (sound power level according to ISO 7779)
• During scanning
max. 65 dB(A)
• Standby
max. 45 dB(A)
Radio frequency emission
according to EN 55011, Class B and Radio
frequency emission FCC, Part 15, Subchapter B,
Class B
Heat emission
• During scanning
max. 1610 W
• Standby
250 W
Cassette return time
51 - 95 secs
Performance
Throughput
71 plates/h (size dependent)
End of Life
Estimated product life
(if regularly serviced and maintained according
to Agfa instructions)
7 yrs.
Preventive maintenance
Preventive maintenance frequency
Note: Preventive maintenance needs to be done
by an Agfa certified field service engineer.
Once a year or after 12000 cycles (whatever
comes first).
Grayscale resolution
Data acquisition
12 bits/pixel
Output to processor
12 bits/pixel
Содержание CR 35-X
Страница 1: ...CR 35 X User Manual 4454C EN 2013 05 14 ...
Страница 4: ...4 4454C EN 2013 05 14 CR 35 X DIGITIZER ...
Страница 40: ...40 4454C EN 2013 05 14 Switching off the CR 35 X CR 35 X DIGITIZER ...
Страница 70: ...70 4454C EN 2013 05 14 Troubleshooting checklist CR 35 X DIGITIZER ...
Страница 71: ...Equipment information sheet Appendix A ...
Страница 77: ...CR cassettes Appendix B ...
Страница 88: ...88 4454C EN 2013 05 14 Technical specifications of the image plates CR 35 X DIGITIZER ...
Страница 89: ...Remarks for HF emission and immunity Appendix C ...
Страница 95: ...95 4454C EN 2013 05 14 Remarks for HF emission and immunity CR 35 X DIGITIZER ...
Страница 96: ...Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 4454C EN 2013 05 14 0413 ...