92
4454C EN 2013-05-14
Remarks for HF-emission and immunity
CR 35-X D
IGITIZER
This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure
that it is used in such an environment.
Tests of Resistance to
Disruption
IEC 60601
Test Level
Level of
Agreement
Electromagnetic Environment
Use portable and mobile radio sets
at a safe distance from the device
(including the leads) not closer
than the recommended protective
distance, which is calculated
according to the equation suitable
for the transmission frequency.
Recommended protective distance:
Conducted high
frequency disturbance
variables in accordance
with IEC 61000-4-6
3 V
eff
150 kHz
to 80 MHz
3 V
eff
d
= 1.2
Radiated high
frequency disturbance
variables in accordance
with IEC 61000-4-3
3 V/m
80 MHz
to 2.5 GHz
3 V/m
d
= 1.2
80 MHz to 800 MHz
d
= 2.3
800 MHz to 2.5 GHz
With
P
as the rated power of the
transmitter in watts (W) in
accordance with the manufacturer
information on the transmitter and
d
as the recommended protective
distance in metres (m).
The field strength of stationary
radio transmitters is lower than the
level of the agreement
a
at all
frequencies in accordance with an
on-site investigation
b
.
Disruptions are possible near
devices that carry the following
symbol:
• REMARK 1: The higher value will apply at 80 MHz and 800 MHz.
• REMARK 2: These Guidelines may not apply to all situations. The dispersion of
electromagnetic waves is influenced by absorption and reflections from buildings,
objects and people.
P
P
P
Содержание CR 35-X
Страница 1: ...CR 35 X User Manual 4454C EN 2013 05 14 ...
Страница 4: ...4 4454C EN 2013 05 14 CR 35 X DIGITIZER ...
Страница 40: ...40 4454C EN 2013 05 14 Switching off the CR 35 X CR 35 X DIGITIZER ...
Страница 70: ...70 4454C EN 2013 05 14 Troubleshooting checklist CR 35 X DIGITIZER ...
Страница 71: ...Equipment information sheet Appendix A ...
Страница 77: ...CR cassettes Appendix B ...
Страница 88: ...88 4454C EN 2013 05 14 Technical specifications of the image plates CR 35 X DIGITIZER ...
Страница 89: ...Remarks for HF emission and immunity Appendix C ...
Страница 95: ...95 4454C EN 2013 05 14 Remarks for HF emission and immunity CR 35 X DIGITIZER ...
Страница 96: ...Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 4454C EN 2013 05 14 0413 ...