53
4454C EN 2013-05-14
Reading an emergency image plate
CR 35-X D
IGITIZER
Reading an emergency image plate in the
dedicated configuration
In the dedicated configuration, you can digitize unidentified emergency
image plates. The digital image data will be assigned default ID data.
To read an emergency image plate:
1
Check that the CR 35-X is ready for operation:
•
The CR 35-X must display the operator main screen with ‘Ready’ status, e.g.:
•
The status indicator at the top of the CR 35-X must be green and be lit constantly.
2
Press the appropriate
emergency button at the front of the
CR 35-X.
The emergency button determines the speed class, i.e. the sensitivity, which will be
used to digitize the image plate.
WARNING:
To digitize an emergency image plate, you must press the
Confirm key after you have inserted the cassette (refer to step
).
Image data transmission to the preview and the image processing
station will be suspended until you have pressed the Confirm key.
Note: The CR 35-X is operational if the status field equals ‘READY’, even if
status messages of the destination are shown (e.g. ‘QS not ready’).
For digitizing images of the trunk.
For digitizing images of the limbs.
QS not ready
READY
Status field
Содержание CR 35-X
Страница 1: ...CR 35 X User Manual 4454C EN 2013 05 14 ...
Страница 4: ...4 4454C EN 2013 05 14 CR 35 X DIGITIZER ...
Страница 40: ...40 4454C EN 2013 05 14 Switching off the CR 35 X CR 35 X DIGITIZER ...
Страница 70: ...70 4454C EN 2013 05 14 Troubleshooting checklist CR 35 X DIGITIZER ...
Страница 71: ...Equipment information sheet Appendix A ...
Страница 77: ...CR cassettes Appendix B ...
Страница 88: ...88 4454C EN 2013 05 14 Technical specifications of the image plates CR 35 X DIGITIZER ...
Страница 89: ...Remarks for HF emission and immunity Appendix C ...
Страница 95: ...95 4454C EN 2013 05 14 Remarks for HF emission and immunity CR 35 X DIGITIZER ...
Страница 96: ...Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 4454C EN 2013 05 14 0413 ...